RecallWatchUSA

Home / Brands / Allen Medical Systems

Allen Medical Systems Recalls

6 recalls on record · March 27, 2013 to April 4, 2018 · Medical Devices

Recalls by year

13
17
18

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

April 4, 2018Medical DevicesFDA
Allen Medical Systems recalls F-T3SYSTEM, T3 (Trumpf Table Top) System The device is a surgical table accessory system used to allow prone positi
Hazard: The recalling firm has become aware that the rail section clamps used to attach the leg support of the device may fail during use, which could result in critical harm to the patient.
February 28, 2018Medical DevicesFDA
Allen Medical Systems recalls Radiolucent Surgery Spine Frame (F70801). Surgical platform used to position patients for lumbar spine procedures.
Hazard: The base of the frame has the potential to crack when the patient is initially transferred to the Bow Frame.
February 28, 2018Medical DevicesFDA
Allen Medical Systems recalls Allen Medical Bow Frame Model A-70800. Surgical platform used to position patients for lumbar spine procedures.
Hazard: The base of the frame has the potential to crack when the patient is initially transferred to the Bow Frame.
February 22, 2017Medical DevicesFDA
Allen Medical Systems, Inc. recalls Arthrex Shoulder Suspension Tower; Catalog Number: AR-1650-01 Product Usage: The Arthrex Lateral Decubitus Shoulder
Hazard: During use, there is a potential scenario which could allow the fastening screws attaching covers to come loose and potentially land in the sterile field causing a delay in surgery.
January 4, 2017Medical DevicesFDA
Allen Medical Systems, Inc. recalls Arthex S3 Arm Sleeve Connector; Catalog Number AR-1650-02; manufactured and distributed between March and November 2016.
Hazard: During use, the unit could separate un-expectantly. This failure would allow the patient's arm to fall out of suspension during a procedure, possibly causing moderate injury to the patient. This potential defect is not apparent to the user.
March 27, 2013Medical DevicesFDA
Allen Medical Systems, Inc. recalls Allen Traction Boot, Modular, OT 1000 Traction Device Boot Assembly, Fracture Table Orthopedic Leg Holders, non-invasiv
Hazard: Some units may have been assembled incorrectly which could result in the boot breaking free from the mounting base and allowing the patient's leg to fall.