Ameditech Recalls
109 recalls on record · October 22, 2014 to October 7, 2020 · Medical Devices
Recalls by year
14
15
17
18
19
20
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls Immutest 5 Pnl Drug Cup w/ Adult -waived, Item No. IMCA-5AB-W
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls Immutest 11 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-11OP
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls Immutest 10 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-10M5
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls iCup DX 8 Panel w/Adulterant, Item No. I-DXA-187-016
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls iCup DX 6 Panel w/Adulterant, Item No. I-DXA-167-013
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls iCup DX 12 Panel w/ Adulterant, Item No. I-DXA-1127-023
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls iCup DX 11 Panel w/ Adulterant, Item No. I-DXA-1117-131
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls iCup DX 10 Panel w/Adulterant, Item No. I-DXA-1107-142
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls EDI BZO Uncut Sheet - Cup*, Item No. 160170
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugSmart Dip AMP/OPI2/COC/THC/BZO Test, Item No. 30520D
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugSmart Dip (MET/OPI2)+AMP+BZO+COC+THC, Item No. 30600D
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugSmart Cup THC/COC/AMP//MET/OPI2/BZO, Item No. 60600D
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugSmart Cup 10, Item No. 61093D-3C, 61085D, 61015D
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugSmart 9 Test Cup (BUP), Item No. 60960D
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugSmart 12 Test Cup (Multi & BUP), Item No. 61210D DrugSmart 12 Test Cup (Multi/Single/BUP), Item No. 61205D
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugSmart 11 Test Cup, Item No. 61127D
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugSmart 10 Test Cup (PCP & OXY), Item No. 61020D
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls DrugCheck Waived 12 Panel Cup w/Adult, Item No. DCC-81205-5
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls CLIA RDTC 12-Panel Cup (w/Adulteration), Item No. CLIA-RDTC-12, DCC-81205-5
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls BZO Strip Dip Card (300ng/ml)*, Item No. 100170
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls 6 Panel Cup Body and Lid w/Adult (AU), Item No. TDDA-6MBAU-CUP
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls 12-Panel Dip Drug Screen with adulterant, Item No. PSDA-12BUP, PSD-6MTDBO300, PSD-10PPX, PSD-10MOX, 60960D, 61127D, DCC-
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls 11 Panel Card (OPI 300) w/Adult, Item No. 71125AA
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls 11 Panel Dip Card (OPI 300) w/Adult, Item No. 11125AA
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls 10-Panel Pipette Drug Screen, Item No. PSP-10M
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
December 30, 2015Medical DevicesFDA
Ameditech Inc recalls 1 Step 12 Panel Cup, Item No. NBCA-12M-W
Hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
November 25, 2015Medical DevicesFDA
Ameditech Inc recalls THERMO FISHER 6 DRUG CASSETTE, Item No. MTPA-6MBAU, MTPA-8MBAU These Drugs of Abuse Tests are one-step immunoassays fo
Hazard: Ameditech is recalling drug abuse tests because they have shown reduced reactivity.
November 25, 2015Medical DevicesFDA
Ameditech Inc recalls The Drug Detection Agency 6 Panel Cup Body and Lid w/Adult (AU), Item No. TDDA-6MBAU-CUP These Drugs of Abuse Tests
Hazard: Ameditech is recalling drug abuse tests because they have shown reduced reactivity.
November 25, 2015Medical DevicesFDA
Ameditech Inc recalls ProScreen 10 Drug Panel Cassette Test, Item No. CPSP-10PPX; ProScreen 10 Panel Cup, Item No. PSCup-10MMO-300, PSCup-10M
Hazard: Ameditech is recalling drug abuse tests because they have shown reduced reactivity.
November 25, 2015Medical DevicesFDA
Ameditech Inc recalls Premier Clia Waived 12 panel Tilt Cup, Item No. C12APC These Drugs of Abuse Tests are one-step immunoassays for the q
Hazard: Ameditech is recalling drug abuse tests because they have shown reduced reactivity.