Aomori Olympus Recalls
113 recalls on record · February 17, 2021 to September 13, 2023 · Medical Devices
Recalls by year
21
23
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 4MMx 25G LOWER, model no. NM-610U-0425 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 4MMx 23G LOWER, model no. NM-610U-0423 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 3MMx 26G LOWER, model no. NM-610U-0326 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 3MMx 23G LOWER, model no. NM-610U-0323 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 6MMx 25G UPPER, model no. NM-610L-0625 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 6MMx 23G UPPER, model no. NM-610L-0623 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 6MMx 21G UPPER, model no. NM-610L-0621 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 5MMx 25G UPPER, model no. NM-610L-0525 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 3MMx 23G LOWER, model no. NM-610L-0523 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 5MMx 21G UPPER, model no. NM-610L-0521 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 4MMx 26G UPPER, model no. NM-610L-0426 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 4MMx 25G UPPER, model no. NM-610L-0425 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 4MMx 23G UPPER, model no. NM-610L-0423 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls NEEDLEMASTER 4MMx 21G UPPER, model no. NM-610L-0421 - Product Usage: Used with endoscopes to perform endoscopic vascular
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX UP 2MM CH 25G 6MM, model no. NM-401L-0625 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX UP 2MM CH 23G 6MM, model no. NM-401L-0623 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX UP 2MM CH 25G 5MM, model no. NM-401L-0525 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX UP 2MM CH 23G 3MM, model no. NM-401L-0523 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX UP 2MM CH 25G 4MM, model no. NM-401L-0425 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX SIF 270CM 23G 4MM, model no. NM-401L-0423 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX SIF 270CM 23G 4MM, model no. NM-400Y-0423 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX LOWER 25G 6MM, model no. NM-400U-0625 - Product Usage: Used with endoscopes to perform endoscopic inje
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX LOWER 25G 5MM, model no. NM-400U-0525 - Product Usage: Used with endoscopes to perform endoscopic inje
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX LOWER 23G 6MM 5/BX, model no. NM-400U-0623 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX LOWER 23G 5MM, model no. NM-400U-0523 - Product Usage: Used with endoscopes to perform endoscopic inje
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX 25G 2.5MM, model no. NM-400U-0425 - Product Usage: Used with endoscopes to perform endoscopic injectio
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX LOWER 23G 4MM, model no. NM-400U-0423 - Product Usage: Used with endoscopes to perform endoscopic inje
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX LOWER 23G 3MM, model no. NM-400U-0323 - Product Usage: Used with endoscopes to perform endoscopic inje
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX UP 2.8MMCH 25G 6MM, model no. NM-400L-0625 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
February 17, 2021Medical DevicesFDA
Aomori Olympus Co., Ltd. recalls INJECTOR FORCEMAX UP 2.8MMCH 25G 5MM, model no. NM-400L-0525 - Product Usage: Used with endoscopes to perform endoscopic
Hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.