ARROW INTERNATIONAL Recalls
525 recalls on record · June 27, 2012 to June 24, 2026 · Medical Devices
Recalls by year
12
13
14
15
16
17
18
19
20
21
22
23
24
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
October 29, 2014Medical DevicesFDA
Arrow International Inc recalls ASK-05501-AFH1, Lot Number 23F13K0969 Anesthesia Conduction Catheter, Epidural Anesthesia Kit
Hazard: Arrow International, Inc. is notifying customers that lot 23F13K0969 of product code ASK-05501-AFH1 contains an incorrect component. The lidstock indicates that the kit contains one Syringe, 10mL, Luer-Slip Loss of Resistance, Glass but contains one Syringe, 10mL, Luer-Lock Loss of Resistance, Glass.
October 22, 2014Medical DevicesFDA
Arrow International Inc recalls Multi-Lumen Central Venous Catheterization Kit
Hazard: Arrow International, Inc. has initiated a voluntary recall for product code AK- 12703, lot number RF3039028 because the label on the outer corrugate is labeled with the incorrect expiration date. All kits packaged within the corrugate are correctly labeled. There is no risk to the patient as the kits within the corrugate are labeled with the correct expiration date.
October 22, 2014Medical DevicesFDA
Arrow International Inc recalls Arrow Select Kits (ASK) and Maximal Barrier Precautions Kits (CDC) with BD Medical Posiflush 10mL Saline Flush Syringe.
Hazard: The component supplier recalled their 0.9% Sodium Chloride Injection USP BD Posiflush SF Saline Flush Syringe 10mL because some unit packages may exhibit open seals which impact package integrity and potentially product sterility.
June 11, 2014Medical DevicesFDA
Arrow International Inc recalls Arrow¿ Percutaneous Sheath Introducer Kits The Percutaneous sheath introducer products permit venous or arterial acce
Hazard: Arrow is recalling the Arrow¿ Percutaneous Sheath Introducer Kits due to dilator tip damage noted during internal inspection. Dilator tip damage has the potential to result in vessel damage.
May 28, 2014Medical DevicesFDA
Arrow International Inc recalls Arrowgard Blue and Arrowgard Blue Plus Pressure Injectable Central Venous Catheter, a multiple-lumen catheter that permi
Hazard: The labeling states that the kit contains Introducer Needle: Echogenic 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe; however, the kit contains, and should state, Introducer Needle: 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe. As a result of this labeling inconsistency, if the clinician were to place the needle under x-r
April 30, 2014Medical DevicesFDA
Arrow International Inc recalls Arrowgard Blue Plus Antimicrobial Multi-lumen Central Venous Catheter, catalogue number AK-45854-CDC.
Hazard: Arrow is recalling the Arrowgard Blue Plus Antimicrobial Multi-lumen Central Venous Catheter due to a labeling inconsistency. One portion of the lidstock states that the kit contains 8.5FR x 16CM catheter, while the rest of the lidstock correctly identifies the catheter as 8.5FR x 20CM
April 30, 2014Medical DevicesFDA
Arrow International Inc recalls Arrow Pressure Injectable PICC Powered by Arrow VPS Stylet Catalogue number CDC-35052-VPS, peripherally inserted central
Hazard: Kits are packaged with the incorrect catheter. The kits contain a 40cm catheter and should contain a 50cm catheter.
December 25, 2013Medical DevicesFDA
Arrow International Inc recalls Continuous Peripheral Nerve Block Catheter Kit and Set Continuous Peripheral Nerve Block Catheter Kit and Set permits
Hazard: A labeling inconsistency was discovered in which the lidstock states that the kit contains a stimulator cable with 1.5mm socket, however, the kit contains a stimulator cable with 1.5mm pin.
December 18, 2013Medical DevicesFDA
Arrow International Inc recalls Arrow Transradial Artery Access products Arrow Transradial Artery Access products are used for percutaneous introduct
Hazard: There is a risk that the introducer needles packaged within the kits can be partially or fully occluded, which can cause delayed treatment or an adverse event when used on a patient.
August 21, 2013Medical DevicesFDA
Arrow International Inc recalls Arrow International, Inc. Two-Lumen CVC Kit with Blue Flex Tip Catheter. The multiple lumen catheter permits venous acce
Hazard: Arrow is recalling the product due to a labeling inconsistency. One portion of the lidstock states the kit contains a 3-Lumen catheter while the rest of the lidstock correctly identifies the catheter as 2-Lumen.
April 17, 2013Medical DevicesFDA
Arrow International Inc recalls Arrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 40 cc; Product Number: IAB-05840-LWS. Th
Hazard: Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
April 17, 2013Medical DevicesFDA
Arrow International Inc recalls Arrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 30 cc; Product Number: IAB-05830-LWS. T
Hazard: Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
April 10, 2013Medical DevicesFDA
Arrow International Inc recalls Arrow/Teleflex Medical Berman Angiographic Balloon Catheter and Reverse Berman Angiographic Balloon Catheter. The Arr
Hazard: The product specifications sheet, part of the Instructions for Use (IFU), for the specified product numbers, reference Renografin-76 contrast medium (viscosity of 8.4 centipoise), which is no longer available. Firm is in the process of updating the labeling for these products. In the meantime, when choosing an alternative contrast medium, health care providers are instructed to select a contrast
July 4, 2012Medical DevicesFDA
Arrow International Inc recalls Arrow Two-Lumen Central Venous Catheterization Set with Blue FlexTip(R) ARROWgard Blue(R) Catheter Product Usage: Th
Hazard: Lidstock of product does not contain chlorhexidine contraindication and contains wording "contains no medication" on the label. Product insert does reflect medicated status.
June 27, 2012Medical DevicesFDA
Arrow International Inc recalls Arrow International, Inc. 7FR 3-Lumen Arrowg+ard Blue Plus(R) Pressure Injectable Central Venous Catheter - Catalog numb
Hazard: Arrow is recalling the product due to the possibility of voids in the catheter hub which could lead to an interlumen crossover.