Home / Brands / Becton, Dickinson and
Becton, Dickinson and Recalls
3 recalls on record · May 27, 2026 to May 27, 2026 · Medical Devices
Also filed under
The recall feeds record what looks like the same company under more than one name:
May 27, 2026Medical DevicesFDA
Becton, Dickinson and Company recalls Swan-Ganz Pacing Catheter, Models: D200F7;
Hazard: Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
May 27, 2026Medical DevicesFDA
Becton, Dickinson and Company recalls Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P
Hazard: Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
May 27, 2026Medical DevicesFDA
Becton, Dickinson and Company recalls Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, C146F7, 83