Becton, Dickinson and Company, BD Bio Sciences Recalls

10 recalls on record · October 9, 2024 to August 13, 2025 · Medical Devices

Recalls by year

24
25

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

August 13, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls BD Trucount Controls, 30 Test - REF: 340335
Hazard: Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.
August 13, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD
Hazard: Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.
July 9, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.
Hazard: Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.
July 9, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls REF: 335775 , BD FACS, 7-Color Setup Beads, 25 setups per kit, Rx Only, IVD. Used in flow cytometry for fluorescence com
Hazard: Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.
January 1, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls FACSLyric 3L12C Instrument RUO, Catalog Number 663518, UDI 00382906635181
Hazard: On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users.
January 1, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls FACSLyric 3L10C Instrument US-IVD, Catalog Number 662878, UDI 00382906628787 The K170974 clearance document describes
Hazard: On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users.
January 1, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls FACSLyric 3L8C Instrument US-IVD, Catalog Number 662877, UDI 00382906628770 The K170974 clearance document describes
Hazard: On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users.
January 1, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls Device Description: FACSLyric 2L6C Instrument US, Catalog Number 662876, UDI 00382906628763 The K170974 clearance
Hazard: On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users.
January 1, 2025Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls FACSLyric 2L4C Instrument US, Catalog Number 662875, UDI 00382906628756 BD FACSLyric" flow cytometer is intended for
Hazard: On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users
October 9, 2024Medical DevicesFDA
Becton, Dickinson and Company, BD Bio Sciences recalls BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), RE
Hazard: Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.