January 21, 2015Medical DevicesFDA
Brainlab AG recalls iPlan RT Dose is a stereotactic radiation treatment planning system that is intended for use in stereotactic, conformal,
Hazard: iPlan RT Radiation Treatment Planning Software: Potentially incorrect patient positioning when using multiple localized CT image data sets.
August 20, 2014Medical DevicesFDA
Brainlab AG recalls Navigation Spine & Trauma 3D Version 2.0 and 2.1 Is intended as an intraoperative image-guided localization system t
Hazard: Brainlab Navigation Software Spine & Trauma 3D 2.0/2.1 offers automatic registration of intraoperatively acquired CT image data sets. To enable automatic registration, the software requires the gantry position of the scanner. The gantry position can either be entered manually or submitted automatically from the CT scanner. If automatic gantry communication is available, the so-called "
August 13, 2014Medical DevicesFDA
Brainlab AG recalls ExacTrac is intended to be used to place patients at an accurately defined point within the treatment beam of a medical
Hazard: When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite displaying the green "OK" icon. If this anomaly occurs and is not detected by the user, the radiation treatment dose at the linear accelerator may be delivered to the unintended target position.
August 6, 2014Medical DevicesFDA
Brainlab AG recalls Patient Data Manager 2.0 (Content manager 2.0, Patient Browser 4.0, DICOM Viewer 2.0) Brainlab Digital Lightbox. A s
Hazard: The following defects have been identified for data loaded with Patient Data Manager version 2.0 (including subversions 2.0.0, 2.0.1 and 2.0.2): 1) When loading non-square pixel image data with PDM 2.0, the data set is not correctly transferred and displayed in the corresponding Brainlab navigation or planning software. When a data set is loaded, the pixels of the data set are sorted in order to
June 11, 2014Medical DevicesFDA
Brainlab AG recalls ExacTrac 5.5, an Image Processing System used for patient positioning for radiation therapy or radiosurgery.
Hazard: When using a specific plan update workflow in combination with a non-Brainlab treatment planning system (TPS), there is a possibility in ExacTrac 5.5.5 to generate an unintended blend of two initially separate plans of the same patient.
May 7, 2014Medical DevicesFDA
Brainlab AG recalls ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIO
Hazard: When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.
April 23, 2014Medical DevicesFDA
Brainlab AG recalls Brainlab, Spine & Trauma 3D 2.0, Navigation Software. An intraoperative image-guided localization system to enable mini
Hazard: The "fluoro match registration" function of the Navigation Software Spine & Trauma 3D 2.0 allows the user to intraoperatively match e.g., CT data sets to the current patient anatomy visible on fluoroscopic images, used by the navigation software during spinal surgeries to display the position and orientation of instruments. For this registration function in combination with a digitally integrated
February 19, 2014Medical DevicesFDA
Brainlab AG recalls BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally in
Hazard: Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect
September 4, 2013Medical DevicesFDA
Brainlab AG recalls BRAINLAB; FRAMELESS SRS QA TARGET POINTER Robotics is a device used to compensate rotational patient misalignment (ro
Hazard: The Frameless SRS QA Target Pointer - Pointer Cap with the engraved cross hairs may become loose even when only a moderate force is applied to the Pointer Cap.