Home / Brands / Change Healthcare Israel
Change Healthcare Israel Recalls
2 recalls on record · June 5, 2019 to June 5, 2019 · Medical Devices
June 5, 2019Medical DevicesFDA
Change Healthcare Israel Ltd. recalls McKesson Cardiology Hemo
Hazard: users are not notified of procedure medication discrepancies between the Vitals and Meds, the Procedure Notes and Patient Common Data screens in Horizon/McKesson Cardiology Hemo" (Hemo). The discrepancy may also affect administered medication data in reports generated from Hemo or the Horizon/McKesson Cardiology Physician s Report as well as customers, who have implemented an outbound interface of procedure results.
June 5, 2019Medical DevicesFDA
Change Healthcare Israel Ltd. recalls Horizon Cardiology Hemo
Hazard: users are not notified of procedure medication discrepancies between the Vitals and Meds, the Procedure Notes and Patient Common Data screens in Horizon/McKesson Cardiology Hemo" (Hemo). The discrepancy may also affect administered medication data in reports generated from Hemo or the Horizon/McKesson Cardiology Physician s Report as well as customers, who have implemented an outbound interface of procedure results.