Home / Brands / Channel Medsystems
Channel Medsystems Recalls
2 recalls on record · May 29, 2024 to April 23, 2025 · Medical Devices
Recalls by year
24
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
April 23, 2025Medical DevicesFDA
Channel Medsystems, Inc. recalls Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-70
Hazard: Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device
May 29, 2024Medical DevicesFDA
Channel Medsystems, Inc. recalls Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-70
Hazard: Error code 003 was not listed in certain rows of the Device Instructions for Use Troubleshooting section (i.e., Table 22. Summary of LCD Messages and Error Codes).