CP Medical Recalls
3 recalls on record · April 26, 2017 to December 18, 2019 · Medical Devices
Recalls by year
17
19
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
December 18, 2019Medical DevicesFDA
CP Medical Inc recalls Coated PGA Undyed Polyglycolic Acid Braided Absorbable Suture Sterile EO 18" (45cm); UDI: (01)10790986000657 (17)220430
Hazard: Failed to meet USP minimum average value for knot pull (tensile) strength.
March 27, 2019Medical DevicesFDA
CP Medical Inc recalls Plain and Chromic Gut Absorbable Surgical Sutures, sterile. 12 sutures per box. Product Usage: Plain Gut Sutures are
Hazard: Potential compromise of product sterility after routine monitoring for Endotoxins showed an out of limit test result.
April 26, 2017Medical DevicesFDA
CP Medical Inc recalls PGCL [poly (glycolide-co-caprolactone)] Absorbable Suture, Manufacturer CP Medical, Portland OR, Sterile, Product Codes
Hazard: The tensile strength minimum as directed by USP <881> was not met over the length of the product's shelf life.