2 recalls on record · June 8, 2022 to August 27, 2025 · Medical Devices
Recalls by year
22
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
August 27, 2025Medical DevicesFDA
Dynex Technologies, Inc. recalls Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with th
Hazard: The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.
June 8, 2022Medical DevicesFDA
Dynex Technologies, Inc. recalls DYNEX Agility, Agility Analyzer, Model No. 67000
Hazard: Control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit will be used instead of the correct control, which may lead to delayed patient results.