Erbe USA Recalls
4 recalls on record · April 1, 2026 to August 19, 2026 · Medical Devices
August 19, 2026Medical DevicesFDA
Erbe USA Inc recalls ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
Hazard: A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
April 1, 2026Medical DevicesFDA
Erbe USA Inc recalls Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use
Hazard: Probes may rupture/burst during activation
April 1, 2026Medical DevicesFDA
Erbe USA Inc recalls Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
Hazard: Probes may rupture/burst during activation
April 1, 2026Medical DevicesFDA
Erbe USA Inc recalls Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use
Hazard: Probes may rupture/burst during activation