July 24, 2024Medical DevicesFDA
Ethicon Sarl, a Johnson & Johnson Company recalls SURGICEL FIBRILLAR 1INX2IN(2.5CMX5.1CM), Product Number: 1963
Hazard: Some foil pouches were found torn open, impacting the sterile barrier which was traced to a specific machine used during product packaging.
July 24, 2024Medical DevicesFDA
Ethicon Sarl, a Johnson & Johnson Company recalls SURGICEL FIBRILLAR 1INX2IN(2.5CMX5.1CM), Product Number: 1961
Hazard: Some foil pouches were found torn open, impacting the sterile barrier which was traced to a specific machine used during product packaging.