ETHICON Recalls
76 recalls on record · July 4, 2012 to October 1, 2025 · Medical Devices
Recalls by year
12
13
14
15
16
18
19
20
21
22
23
24
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
April 18, 2018Medical DevicesFDA
Ethicon, Inc. recalls PDS Plus Antibacterial (Polydioxanone) Suture Size 0 PDS Plus Violet 36 Single Armed CTB-1 Needle
Hazard: Three (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.
April 18, 2018Medical DevicesFDA
Ethicon, Inc. recalls PDS Plus Antibacterial (Polydioxanone) Suture Size 1 PDS PLUS Violet 27 Single Armed OS-4 Needle
Hazard: Three (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.
April 18, 2018Medical DevicesFDA
Ethicon, Inc. recalls PDS Plus Antibacterial (Polydioxanone) Suture Size 0 PDS Plus Violet 36 Single Armed CT-1 Needle
Hazard: Three (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.
March 28, 2018Medical DevicesFDA
Ethicon, Inc. recalls DERMABOND" PRINEO" Skin Closure System Product Usage: DERMABOND PRINEO System is intended for topical application
Hazard: Ethicon discovered that specific lots of DERMABOND"PRINEO"System may not dry within the specified time after proper application, and thus may fall off.
March 7, 2018Medical DevicesFDA
Ethicon, Inc. recalls Coated VICRYL RAPIDETM (Polyglactin 910) Suture
Hazard: On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture damage was identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent investigation, the suture also failed knot tensile strength testing.
October 5, 2016Medical DevicesFDA
Ethicon, Inc. recalls ETHICON MERSILENE" Tape (also referred to as MERSILENETM Polyester Fiber Strip or MERSILENETM Fiber Ligature)
Hazard: ETHICON MERSILENE" Tape (also referred to as MERSILENETM Polyester Fiber Strip or MERSILENETM Fiber Ligature) products are being recalled since a group of MERSILENE" Tape product codes are being supplied with an incorrect Instructions for Use (IFU) insert.
June 22, 2016Medical DevicesFDA
Ethicon, Inc. recalls EVARREST Fibrin Sealant Patch, Manufactured by: Omrix Biopharmaceuticals, Ltd, Israel. Indicated for use with manual
Hazard: Ethicon is recalling EVARREST Fibrin Sealant Patch because the characters "0-bad" were printed in the expiration date field on the foil pouch labels of the Fibrin Patch units instead of the valid expiration date (2016-12-28).
January 13, 2016Medical DevicesFDA
Ethicon, Inc. recalls Gynecare Thermachoice III Uterine Balloon Therapy System Ethicon, Inc. The GYNECARE THERMACHOICE¿ III Uterine Balloon
Hazard: Stability data does not substantiate the labeled two-year shelf life of affected product.
July 1, 2015Medical DevicesFDA
Ethicon, Inc. recalls Ethicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm)
Hazard: Ethicon Fast Absorbing Surgical Gut (Plain) Suture is being recalled because the labeling on the box and foil are incomplete.
December 10, 2014Medical DevicesFDA
Ethicon, Inc. recalls Ethicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Hazard: Ethicon is unable to ensure that PERMA-HAND silk suture (siliconized black braid) meets the expected sterility assurance level for sterile products.
April 9, 2014Medical DevicesFDA
Ethicon, Inc. recalls Proceed Surgical Mesh, Product Codes PCDB1, PCDT1, and PCDJ1. Ethicon Inc., Johnson & Johnson Medical GmbH, used f
Hazard: The possibility exists that an incomplete seal on the packaging compromised the sterility of the device and introduced the potential for delamination.
December 18, 2013Medical DevicesFDA
Ethicon, Inc. recalls Ethicon Stratafix Symmetric PDS Plus Knotless Tissue Control Device Ethicon LLC Guaynabo, Puerto Rico 00969. Intended
Hazard: Ethicon Stratafix Symmetric PDS Plus Knotless Tissue Control Device has a small number of tab failures and fascial dehiscences in lower abdominal incisions.
November 28, 2012Medical DevicesFDA
Ethicon, Inc. recalls ETHICON SURGIFLO HEMOSTATIC MATRIX Distributed by ETHICON ETHICON, Inc. Somerville, NJ 08876, USA SURGIFLO Hemostat
Hazard: SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.
November 28, 2012Medical DevicesFDA
Ethicon, Inc. recalls ETHICON SURGIFLO HEMOSTATIC MATRIX KIT Plus FlexTip with Thrombin distributed by ETHICON ETHICON Inc. Somerville, NJ 08
Hazard: SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.
October 31, 2012Medical DevicesFDA
Ethicon, Inc. recalls EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital,
Hazard: Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be related to the use of the spray device at higher-than-recommended pressures and/or in close proximity to the tissue surface. The field correction initiated is to provide the recommend
July 4, 2012Medical DevicesFDA
Ethicon, Inc. recalls Ethicon Surgiflo Hemostatic Matrix Kit with Thrombin Plus Flex Tip Distributed by Ethicon Ethicon Inc., Somerville, NJ
Hazard: The lyophilized thrombin component of the Surgiflo Hemostatic Matrix kit with Thrombin has the potential to be non-sterile due to an undetected air leak which occurred during the manufacturing process.