2 recalls on record · December 5, 2018 to December 5, 2018 · Medical Devices
December 5, 2018Medical DevicesFDA
EUROTROL INC recalls HemoTrol Level 3- In Vitro Diagnostic Hematology quality control mixture. Catalogue number: 171.003.002 Eurotrol Hem
Hazard: Incorrect measurement results caused by microbial contamination of the product, results in values lower than assigned values
December 5, 2018Medical DevicesFDA
EUROTROL INC recalls HemoTrol Level 2- In Vitro Diagnostic Hematology quality control mixture Catalogue number: 171.002.002 Eurotrol Hemo
Hazard: Incorrect measurement results caused by microbial contamination of the product, results in values lower than assigned values