Flowonix Medical Recalls
13 recalls on record · July 26, 2017 to July 7, 2021 · Medical Devices
Recalls by year
17
19
20
21
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
July 7, 2021Medical DevicesFDA
Flowonix Medical Inc recalls Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable In
Hazard: Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection
February 19, 2020Medical DevicesFDA
Flowonix Medical Inc recalls Prometra II Programmable 20mL Pump. Catalog No. 93827 The device is an implantable infusion pump. GTIN-14 00810335020YY
Hazard: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
February 19, 2020Medical DevicesFDA
Flowonix Medical Inc recalls Prometra Programmable 20mL Pump. Catalog No. 91827 The device is an implantable infusion pump. GTIN-14 00810335020XXX (X
Hazard: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
February 19, 2020Medical DevicesFDA
Flowonix Medical Inc recalls US Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14 00810335020082
Hazard: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
February 19, 2020Medical DevicesFDA
Flowonix Medical Inc recalls Prometra II 40mL Pump. Catalog No. 16827 The device is an implantable infusion pump. GTIN-14 00810335020242
Hazard: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
February 19, 2020Medical DevicesFDA
Flowonix Medical Inc recalls Prometra II 20mL Pump. Catalog No. 13827 The device is an implantable infusion pump. GTIN-14 00810335020228
Hazard: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
February 19, 2020Medical DevicesFDA
Flowonix Medical Inc recalls Prometra Programmable 20mL Pump. Catalog No. 11827 The device is an implantable infusion pump. GTIN-14 00810335020082
Hazard: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
June 12, 2019Medical DevicesFDA
Flowonix Medical Inc recalls Prometra II Programmable Pump (Clinical), Cat. No. 01827 For intrathecal infusion.
Hazard: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.
June 12, 2019Medical DevicesFDA
Flowonix Medical Inc recalls Prometra II Programmable Pump, Cat. No. 13827 For intrathecal infusion.
Hazard: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.
June 12, 2019Medical DevicesFDA
Flowonix Medical Inc recalls Prometra Programmable Pump, Cat. No. 11827 For intrathecal infusion.
Hazard: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.
June 12, 2019Medical DevicesFDA
Flowonix Medical Inc recalls Prometra Programmer (Grand Prime) Software version 2.00.29, Cat. No. 13828, used with Prometra and Prometra II Programma
Hazard: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.
June 12, 2019Medical DevicesFDA
Flowonix Medical Inc recalls Prometra Clinician Programmer (Touchscreen) Software version 2.00.29, Cat. No. 12828, used with Prometra and Prometra II
Hazard: The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.
July 26, 2017Medical DevicesFDA
Flowonix Medical, Inc. recalls PROMETRA II Programmable Infusion Pump, Implantable Infusion Pump, Catalog Number/REF 13827, Part Number PL-31790-02. M
Hazard: Flowonix Medical received a report of a patient implanted with the Prometra II Programmable Pump who may have received a fatal drug overdose during an MRI procedure. The Prometra II pump has an FDA-approved design feature intended to permit safe exposure to an MRI without removing drug from the reservoir.