RecallWatchUSA

Home / Brands / Focus Diagnostics

Focus Diagnostics Recalls

11 recalls on record · November 7, 2012 to April 5, 2017 · Medical Devices

Recalls by year

12
14
15
16
17

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

April 5, 2017Medical DevicesFDA
Focus Diagnostics Inc recalls Legionella IFA Substrate Slide (IF0951) intended for qualitatively detecting and semi-quantitating human serum antibodie
Hazard: After a submission for CLIA database update from Focus Diagnostics to DiaSorin Molecular,the company found that their Legionella IFA device (IF0950) is not 510(k) exempt.
April 5, 2017Medical DevicesFDA
Focus Diagnostics Inc recalls Legionella IFA (IF0950) intended for qualitatively detecting and semi-quantitating human serum antibodies to Legionella
Hazard: After a submission for CLIA database update from Focus Diagnostics to DiaSorin Molecular,the company found that their Legionella IFA device (IF0950) is not 510(k) exempt.
April 13, 2016Medical DevicesFDA
Focus Diagnostics Inc recalls Simplexa Group A Strep Direct (MOL2850) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL
Hazard: Focus Diagnostics is recalling the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850) because it may result in a false negative result, a false positive result or Error code (s) 500 or 505, and it may result in higher invalid rates.
April 13, 2016Medical DevicesFDA
Focus Diagnostics Inc recalls Simplexa Flu A/B & RSV Direct (MOL2650) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL
Hazard: Focus Diagnostics is recalling the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850) because it may result in a false negative result, a false positive result or Error code (s) 500 or 505, and it may result in higher invalid rates.
April 13, 2016Medical DevicesFDA
Focus Diagnostics Inc recalls Simplexa HSV 1&2 Direct (MOL2150) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2150, M
Hazard: Focus Diagnostics is recalling the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850) because it may result in a false negative result, a false positive result or Error code (s) 500 or 505, and it may result in higher invalid rates.
November 18, 2015Medical DevicesFDA
Focus Diagnostics Inc recalls Focus Diagnostics Anaplasma Phagocytophilum IFA IgM Test Kit, Model No. IF1450M in vitro diagnostic.
Hazard: Focus Diagnostics is recalling the Anaplasma phagocytophilum IFA IgM kit because of the potential increase in invalid results.
April 29, 2015Medical DevicesFDA
Focus Diagnostics Inc recalls STRATIFY JCV DxSelect, Model No. EL1950. Assay for detection of antibodies to JC Virus in human serum or plasma. In Vi
Hazard: Focus Diagnostics is recalling the Stratify JCV Dx Select due to the use of a non-conforming batch in the manufacture of JCV DxSelect EL1950 kit lot #27333. This could lead to a potential for false positive JCV results.
July 9, 2014Medical DevicesFDA
Focus Diagnostics Inc recalls Simplexa" Flu A/B & RSV Direct, Model Number: MOL2650. 510(k) K120413  Simplexa" Flu A/B & RSV Direct. The Foc
Hazard: Focus Diagnostics is providing an urgent safety notice for a correction to the labeling for Simplexa" Flu A/B & RSV Direct (MOL2650). Focus Diagnostics received some customer complaints of Simplexa Flu A/B & RSV Direct assays with sporadic false signals, which may result in a higher false results rate due to a potential software spectral matrix and Direct Amplification Disc storage.
March 12, 2014Medical DevicesFDA
Focus Diagnostics Inc recalls Simplexa Flu AlB & RSV Direct Assay Kits, intended for use as an aid in the differential diagnosis of influenza A, influ
Hazard: Focus Diagnostics initiated the recall of the certain Simplexa Flu AlB & RSV Direct assay kits due to the potential of false positive Flu A, Flu B and RSV signals.
February 5, 2014Medical DevicesFDA
Focus Diagnostics Inc recalls Simplexa Flu A/B & RSV Direct assay, Model MOL2650. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is inte
Hazard: Focus Diagnostics is initiating an urgent safety notice correction for Simplexa Flu A/B & RSV Direct assay (MOL2650) because Focus Diagnostics has received some customer complaints of Simplexa Flu A/B & RSV lots due to sporadic false positive RSV signals, which may result in a higher RSV false positive rate.
November 7, 2012Medical DevicesFDA
Focus Diagnostics Inc recalls West Nile Virus IgG DxSelect" kit Catalog No. EL0300G Kit Lots 122150, 122155, 122641, 122650 and 122859 that includes
Hazard: The firm recalled due to higher reactivity with some samples in the effected lots, which may reflect a higher false positive rate.