July 30, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, CARESCAPE Monitor B850, B650 or B450. Patient Monitor.
Hazard: Partial loss of monitoring with the CARESCAPE Monitor B850, B650 or B450. When connecting the PDM (Patient Data Module) to the CARESCAPE B850, B650 or B450 monitor with software version 2 the PDM may not start communicating to the monitor and loss of the PDM parameters could occur.
July 30, 2014Medical DevicesFDA
GE Healthcare, LLC recalls DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES. DATEX-OHMEDA S/5TM FM WITH L-FICU04 ANDL-FICUO
Hazard: Failure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value may impair clinical decision making for both mechanically and spontaneously ventilated patients because of incorrect low CO2 values.
July 9, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Revolution XR/d. Revolution XR/d Intended Use: The Revolution XR/d Digital Radiographic imaging syste
Hazard: GE Healthcare has become aware of a potential safety issue due to a collimator installation error during a service maintenance activity involving the Proteus XR/a and Revolution XR/d X-ray imaging systems. . There was a reported incident of a collimator fall which caused a serious patient injury.
July 9, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Proteus XR/a. Proteus XR/a Intended Use: Is intended for use in generating radiographic images of huma
Hazard: GE Healthcare has become aware of a potential safety issue due to a collimator installation error during a service maintenance activity involving the Proteus XR/a and Revolution XR/d X-ray imaging systems. . There was a reported incident of a collimator fall which caused a serious patient injury.
June 25, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Compact Airway Modules E-CO, E-COV, E-COVX, E-CAiO, E-CAiOV and E-CAiOVX. The Datex-Ohmeda S/5MT Compa
Hazard: Potential loss of airway gas measurement in the Compact Airway Gas Modules. Loosening of the gas sampling pump hex screw in the module pump unit may lead to pump failure that may cause loss of respiratory airway gas parameter monitoring. If this issue occurs, the monitor will show an alarm note on monitor screen of either "Low gas sample flow" or "Check sample gas out".
June 11, 2014Medical DevicesFDA
GE Healthcare, LLC recalls Kenex Radiation Shield and surgical lamps installed with GE Healthcare Interventional, Surgery, CT and X-Ray Imaging sy
Hazard: GE Healthcare has recently become aware of a potential safety issue due to the improper installation of a Kenex ceiling suspended radiation shield and surgical lamp. This issue is not related to the design of the Kenex product. The affected products are limited to installations completed or contracted by GE Healthcare. There has been a reported incident of a radiation shield falling from the ceil
June 4, 2014Medical DevicesFDA
GE Healthcare recalls The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by t
Hazard: It has been identified that in a reject image workflow where multiple studies are simultaneously opened on the Centricity PACS RA1000 workstation and images are rejected from the active displayed study, images for the previously displayed study may also be rejected. The issue can occur with a previously displayed study that remains open in the background. If one or more, but not all, images in
May 28, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01 Product Usage: The Datex-Ohmeda S/5TM E-NMT module is ind
Hazard: When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of the hand are seen after TOF (Train of Four) stimulation, but the patient monitor shows no counts, or counts are not corresponding to the actual a
May 14, 2014Medical DevicesFDA
GE Healthcare recalls Nellcor SpO2 Upgrade kit, consists of a Nellcor SpO2 circuit board, an OEM part manufactured by a third party supplier;
Hazard: Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
May 14, 2014Medical DevicesFDA
GE Healthcare recalls Panda¿ Wall-Mount Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable t
Hazard: Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
May 14, 2014Medical DevicesFDA
GE Healthcare recalls Panda¿ Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable
Hazard: Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
May 14, 2014Medical DevicesFDA
GE Healthcare recalls Panda¿ iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to therm
Hazard: Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
May 14, 2014Medical DevicesFDA
GE Healthcare recalls Giraffe¿ Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-r
Hazard: Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
May 7, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Discovery MR450, Discovery MR750, Optima MR450w, Optima MR450w equipped with the GEM option, Discovery MR7
Hazard: GE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products. Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block mechanism malfunctions or becomes misaligned. This could make the safe removal of the patient from the system difficult.
May 7, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, CardiacVX [CardiacVX (1.1.0 Ext. d) and CardiacVX Flow (1.1.0 Ext. d) analysis software running on Advant
Hazard: GE Healthcare has recently become aware of a potential safety issue due to higher than actual flow values reported on CardiacVX and CardiacVX Flow running on Advantage Workstation.
April 30, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Vivid E9 Ultrasound System, a general-purpose ultrasound system, specialized for use in cardiac imaging.
Hazard: GE Healthcare has recently become aware of a potential safety issue where the probe power surveillance settings for 4 probes (C1-5-D, C2-9-D, IC5-9-D and M5Sc-D) are incorrect. This may result in a situation where probes can overheat.
April 30, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Discovery MR450, Discovery MR750, Optima MR450w, Optima MR 450w GEM, Discovery MR750w, and Discovery MR75
Hazard: Potential safety issue with gradient coil electromechanical connections associated with GE MR Products. Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance can lead to increasing heat during scanning which, in turn, can cause material in the Rear Endbell enclosure to overheat, generate smoke in the scan room, and cause some localize
March 26, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Optima CT580/Discovery CT590RT scanners running software version 13BW16.3. The Discovery CT 590 RT and
Hazard: GE Healthcare has recently become aware of a potential safety issue due to a software issue associated with the Head Scan Protocols used on your Optima CT580 or Discovery CT590RT scanner. A potential hazardous situation may occur during a CT head scan causing a novel and rare artifact that may hide pathology. When using the head and large head SFOV, which applies the Iterative Bone Option (IBO)
March 19, 2014Medical DevicesFDA
GE Healthcare, LLC recalls Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usage: The intended use of the Qua
Hazard: Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home position and then slowly starting drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. N
March 19, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Varicam/VG and VG Hawkeye devices Product Usage: The intended use of the Quasar system is to perform gener
Hazard: Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home position and then slowly starting drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. N
March 19, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Infinia and Infinia Hawkeye Nuclear Medicine Systems. For Dual-Head Variable Angle Gamma Camera, Hawkey
Hazard: Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home position and then slowly starting drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. N
March 19, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare CARESCAPE Monitor B650 and CARESCAPE Monitor B850. K102239 CARESCAPE Monitor B650 The CARESCAPE Monitor
Hazard: GE Healthcare has recently become aware of potential safety issues associated with the CARESCAPE Monitor B850 and CARESCAPE Monitor B650. These issues include both NIBP (Non-Invasive Blood Pressure) and Bed to Bed Issues.
March 19, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova 4100IQ Innova 2121IQ, Innova 3131IQ; Cardi
Hazard: GE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of real-time interventional imaging. No injuries have been reported due to this issue.
March 12, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Model Number/ Description: 5342114 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-2 GSI Viewer
Hazard: When performing a secondary capture generated from GSI Viewer, the header information is correct on the images. However, when changing the ROI (region of interest) and zooming in, the DICOM header does not change to reflect the new image, thus the line measurements may be inaccurate.
March 12, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Solar 8000M and Solar 8000i Patient Monitor software version 5.4, 5.5, or 5.6 with the Automatic-View-On-
Hazard: GE Healthcare has recently become aware of a potential safety issue with the Solar 8000M/i Patient Monitor Automatic-View-On-Alarm (AVOA) feature. The Solar 8000M/i AVOA feature will stop functioning if the patient monitor is discharged from the CIC Central Station while a remote automatic bedside alarm view is currently being displayed on the Solar. Once this issue occurs, subsequent remote AV
March 5, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Precision 500D and Proteus XR/a equipped with Wireless Digital Radiography Option (WDR1). The Wireless
Hazard: GE Healthcare has recently become aware of a software issue associated with patient selection from the worklist on your WDR1(Precision 500D and Proteus XR/a equipped with Wireless Digital Radiography Option) that may impact patient safety. Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the WDR1 are not properly displayed. I
March 5, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare - Giraffe¿ and Panda¿ iRes Infant Warmer Product Usage: The infant radiant warmers provide infrared
Hazard: Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
March 5, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare - Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant Resuscitation System.
Hazard: Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
March 5, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare - Giraffe Warmer;Panda Warmer;Panda iRes Warmer, Giraffe Stand-alone Infant Resuscitation System. Prod
Hazard: Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
February 26, 2014Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Patient Monitors, Dash Port2 Docking Station.
Hazard: GE has become aware of multiple issues affecting the Dash 3000/4000/5000 Patient Monitor and Dash Port 2 Docking Station: DASH: 1. Dash stops communicating on the Unity Network resulting in Loss of Monitoring at the CARESCAPE" CIC Pro Central Station*. 2. Dash displays distorted ECG waveforms in combination monitoring mode. 3. Dash incorrectly becomes Time Master on the Unity Network¿. 4.