October 9, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Discovery MR750 3.0T, Optima MR450w 1.5T, and Discovery MR450 1.5T . scanner designed to support high
Hazard: GE Healthcare has recently become aware of a potential safety issue due to exposed yellow foam on the end bells of Discovery MR450, Discovery MR750, and Optima MR450w products. The yellow foam strip used between the end bell covers and patient bore of certain GE MR Products may extend past the surface of the covers. The foam itself is not hazardous. However, any fluid or body substance absorbed
October 2, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE 1.5T Signa HDxt, 3.0T Signa HDxt, Signa Vibrant, HD23 Software and the following model numbers: 2395001, 2395001-2, 2
Hazard: GE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products. For ClariView and Screen Save from GVTK (Gems Visualization ToolKit) Viewer, the MR Product can generate image headers with non-unique series identifiers, which when exported to some PACS systems, results in the series being associated wit
September 25, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Seno Advantage, 5183685, 5183685-2, 5183685-3, 5183685-4, 5183685-5, 5183685-6, 5183685-7, 5183685-8, 518
Hazard: GE Healthcare has recently become aware that patient mismatch without a warning message could occur on Seno Advantage (SA) workstations under certain circumstances. This issue may impact patient safety.
September 25, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, CARESCAPE Monitor B650. The CARESCAPE Monitor B650 is a multi-parameter patient monitor intended for u
Hazard: GE Healthcare has recently become aware of a number of potential issues associated with the CARESCAPE Monitor B650. No injuries or illnesses have been reported.
September 25, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Discovery IGS 730 Cardiovascular X-Ray Imaging System. The angiography X-ray systems are indicated for
Hazard: GE Healthcare has recently become aware of a potential safety issue with respect to Discovery IGS 730 Imaging Systems. A patient data management error may prevent the Discovery IGS 730 Imaging System from booting on start-up or re-booting during procedure. The boot issue may occur when a large amount of patient data remains in the browser and is not deleted. The inability to boot the system ma
September 25, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Advantage Workstation, READY View Application. Image analysis software.
Hazard: GE Healthcare has become aware that when analyzing fMRI images using the READY View application software, incorrect fMRI activation maps may be computed and displayed.
September 25, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, CARESCAPE Monitor B850. Intended for use in multiple areas within a professional healthcare facility.
Hazard: GE Healthcare has recently become aware of an issue associated with the CARESCAPE Monitor B850 when used with the iCentral network. The iCentral must have the alarm management software license (L-NETM-00) and the alarm limit change option enabled to experience this issue. When changing the HR, SpO2 or Invasive pressure alarm limits from the iCentral over the network, the alarm priority on th
September 18, 2013Medical DevicesFDA
GE Healthcare, LLC recalls Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD V
Hazard: GE Healthcare has recently become aware of a potential safety issue involving the pre-operative check, low pressure leak test at 1% dial setting of the Tec 6 and Tec 6 Plus Vaporizers. The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. If the leak is not detected, it will result in fresh gas leaking to
September 4, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide u
Hazard: GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module. ECG and Respiration signal noise caused by DASH, PDM, and/or Tram ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.
September 4, 2013Medical DevicesFDA
GE Healthcare, LLC recalls SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE (Marketed As GE Healthcare CARESCAPE Patient Data Module) Prod
Hazard: GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module. ECG and Respiration signal noise caused by DASH, PDM, and/or Tram ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.
September 4, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Dash 3000, Dash 4000 and Dash 5000. Product Usage: The intended use of the system is to monitor physi
Hazard: GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module. ECG and Respiration signal noise caused by DASH, PDM, and/or Tram ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.
September 4, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Ultrasound ViewPoint 6. Product Usage: ViewPoint is intended to be used in medical practices and in clinical de
Hazard: GE has recently become aware of a potential safety issue due to the incorrect calculation of the pulsatility index value (PIV) for venous flow measurements associated with obstetric reporting of the GE ViewPoint 6 system. The ViewPoint calculated Doppler PIV for veins could be incorrect. An incorrect calculation of the ductus venosus PIV could lead to false high risk for trisomy 13, 18, and 21,
August 28, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Precision 500D Product Usage: The Precision 500D is designed to perform radiographic and fluoroscopic
Hazard: GE Healthcare has recently become aware of a potential safety issue involving the overhead Video Monitor Suspension involving Precision 500D Systems. There is a potential for the locking nut to loosen out of the Video Monitor Suspension yoke which could result in impact. This will not lead to an immediate fall of the suspended monitor since it will be held in place by the remaining bushing and
August 28, 2013Medical DevicesFDA
GE Healthcare, LLC recalls The Discovery NM/CT 570c CT Standalone Scan The Discovery NM/CT 570c hybrid system is a combination of a cardiac ded
Hazard: GE Healthcare has recently become aware of an incident where the wrong part of the patient body was scanned during a CT standalone exam with the Discovery NM/CT 570c
August 21, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, LOGIQ S8 Ultrasound System The device is intended for use by a qualified physician for ultrasound eva
Hazard: GE Healthcare became aware of a potential issue related to incorrect image measurement associated with the software of the LOGIQ S8 system. This issue was found internally. Users may notice that the system shows incorrect image measurement results on Pulsed Wave (PW) or Time Intensity Curve (TIC) or depth measurement when using Virtual Convex mode and putting a cursor (Sample Volume, Depth measur
August 21, 2013Medical DevicesFDA
GE Healthcare, LLC recalls SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM GE SIGNA 3.0T
Hazard: Screen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system, internal investigation revealed that the issue also impacts multiple GE healthcare MR Systems.
August 21, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Optima XR220amx, Optima XR200amx, and Brivo XR285amx. The series are indicated for use on adult and pe
Hazard: GE Healthcare conducted a recall on various mobile x-ray units (Optima XR220amx, Optima XR200amx, and Brivo XR285amx systems) because of an Erratic Drive Performance.
August 21, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE LightSpeed CT750 HD The GE LightSpeed CT750 HD Computed Tomography X-ray system is intended to produce cross-secti
Hazard: GE Healthcare conducted a recall involving 2 events affecting the GE LightSpeed CT750 HD. Event 1 - There was a report regarding CT number variance in time line at posterior fossa region of head perfusion. When beam hardening and helical artifacts in this region are reconstructed into the perfusion map, these artifacts will give subtle differences in the CT numbers. Event 2 - The software does no
August 7, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, LOGIQ e Ultrasound System. The LOGIQ i/e and Vivid e is a general purpose ultrasound system intended
Hazard: GE Healthcare became aware of a potential issue associated with the LOGIQ e system. Image data from one patient has the potential to be linked to the patient ID of another patient.
August 7, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, VOLUSON E6, E8, E8 EXPERT, E10 DIAGNOSTIC ULTRASOUND SYSTEM. The device is a general purpose ultrasoun
Hazard: GE Healthcare has recently become aware that during use of the Voluson E6 or E8 systems, that an error message Out of Memory can be displayed which could cause the system to freeze. This is an intermittent behavior that also requires a certain workflow. This issue can be resolved by rebooting the system.
July 31, 2013Medical DevicesFDA
GE Healthcare, LLC recalls Optima NM/CT 640 Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging pr
Hazard: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
July 31, 2013Medical DevicesFDA
GE Healthcare, LLC recalls Discovery NM/CT 670 dual detector free-geometry integrated nuclear imaging system. Product Usage: The GE Discover
Hazard: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
July 31, 2013Medical DevicesFDA
GE Healthcare, LLC recalls Discovery NM 630, dual detector free-geometry integrated nuclear imaging system. Product Usage: The GE Discovery N
Hazard: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
July 31, 2013Medical DevicesFDA
GE Healthcare, LLC recalls Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Product Usage: The Brivo NM 615
Hazard: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
July 31, 2013Medical DevicesFDA
GE Healthcare, LLC recalls Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usage: The intended use of th
Hazard: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.
July 31, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Varicam/VG and VG Hawkeye devices Product Usage: The intended use of the Quasar system is to perform g
Hazard: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.
July 31, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Infinia Hawkeye 4 Option For Dual-Head Variable Angle Gamma Camera, Hawkeye Option for Dual-Head Variable
Hazard: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.
July 10, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare, Innova 3100 and Innova 4100 systems. The Digital Fluoroscopic Imaging System is indicated for use in d
Hazard: GE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.
July 3, 2013Medical DevicesFDA
GE Healthcare, LLC recalls CARESCAPE Monitor B850 Product Usage: The CARESCAPE TM Monitor B850 is a multi-parameter high acuity patient monit
Hazard: GE Healthcare become aware of 4 potential issues associated with the CARESCAPE Monitor B850. These issues include CARESCAPE Monitor B850 uDOM failure, CARESCAPE Monitor B850 unstable USB communication, CARESCAPE Monitor B850 remote printing issue with currently printing PRN 50 recorder and CARESCAPE Monitor B850 reboots due to an X server freeze.
July 3, 2013Medical DevicesFDA
GE Healthcare, LLC recalls GE Healthcare Senographe Essential System. The Senographe 2000D system generates digital mamnmographic images that ca
Hazard: GE Healthcare has recently become aware of a potential issue due to inaccessible Operator Instructions associated with the defective Operator Manual CD of the Senographe Essential.