RecallWatchUSA

Home / Brands / Hitachi Medical Systems America

Hitachi Medical Systems America Recalls

17 recalls on record · July 25, 2012 to March 31, 2021 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
21

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

March 31, 2021Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained personnel (doctor, Sonographer, et
Hazard: Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
February 26, 2020Medical DevicesFDA
Hitachi Medical Systems America Inc recalls UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is intended to be used for diagnostic ul
Hazard: These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.
November 13, 2019Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Scenaria Whole-body X-ray CT System
Hazard: There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.
August 29, 2018Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Scenaria CT system Product Usage: The SCENARIA CT system is indicated for head, whole body, cardiac and vasc
Hazard: The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table failed to stop when the control was released and contacted the CT gantry. The problem could not be duplicated at the site, so to avoid further issues while a root cause investigation was pursued, Hitachi choose to add an optional touch sensor to the gantry to add a redundant level of protection. It is conceivable that a patient s extremity may be caught between the tabletop and the gantry surface during uncontrolled motion or that the table contact of the gantry might cause internal damage to the unit.
July 25, 2018Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Oasis MRI system Product Usage: Hitachi MR system is an imaging device and is intended to provide the physic
Hazard: The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt.
March 7, 2018Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Ultrasound Transducer, Model Number: UST-5550-R, is used in conjunction with a diagnostic ultrasound system evaluation d
Hazard: The ultrasound probe may not have adequate protection against electrical shock hazards.
May 24, 2017Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Arietta 70 Ultrasound System Intended for use by trained personnel (doctor, sonographer. etc.) for the diagnostic Ult
Hazard: The strength of the arm mounting mechanism was insufficient and had to be redesigned.
March 29, 2017Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Echelon MRI System
Hazard: Image data transferred from the MRI system to a workstation showed errors on the slice position reference image.
March 29, 2017Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Echelon Oval MRI System
Hazard: Image data transferred from the MRI system to a workstation showed errors on the slice position reference image.
March 29, 2017Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Oasis MRI System
Hazard: Image data transferred from the MRI system to a workstation showed errors on the slice position reference image.
February 1, 2017Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Oasis MRI System - C-Spine Coil
Hazard: The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.
August 17, 2016Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087
Hazard: The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
August 19, 2015Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Oasis Coil Extension Cable used with Hitachi Oasis MRI System Product Usage: The Oasis MR system is an imaging devic
Hazard: The firm received a complaint stating the Coil Extension Cable accessory would overheat and be a burn hazard if touched.
September 10, 2014Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the ph
Hazard: The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard.
June 26, 2013Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR
Hazard: Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.
August 29, 2012Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Echelon MRI System/Oasis MRI, Software. These are software controlled MRI devices. Product Usage: The Echelo
Hazard: Hitachi discovered a software error with a feature called Multi-Planar Reconstruction (MPR). MPR takes an original volume dataset reconstructed in any plane view and creates a new image set from another view (i.e. axial to coronal, coronal to sagittal, etc.). The error causes the first image in the dataset to have a right/left reversed orientation. The software error can occur randomly when execu
July 25, 2012Medical DevicesFDA
Hitachi Medical Systems America Inc recalls Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the
Hazard: Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.
Hitachi Medical Systems America Recalls | RecallWatchUSA