Home / Brands / Human Care USA
Human Care USA Recalls
2 recalls on record · April 6, 2022 to April 6, 2022 · Medical Devices
April 6, 2022Medical DevicesFDA
Human Care USA, Inc. recalls Hanger-Bar 2P 45cm
Hazard: The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.
April 6, 2022Medical DevicesFDA
Human Care USA, Inc. recalls Hanger-Bar 2P 45cm
Hazard: The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.