December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 26 IN(66 cm) APPX 3.4ml, 10 DROP BLOOD SET, 170 MICRON FILTER, Spiros, REF: Z2795
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls Appx 1.7 ml Bag Spike w/Integrated Clave, 2 Drop-In Spiros, REF: CH3260
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls Spinning Spiros Closed Male Luer, Red Cap, REF: IB-CH2000SC; 5" (13 cm) Appx 1.5 ml, Add-On Set w/Spiros, REF: IB-CH3034
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 31" (78 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3011; 40"
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrat
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 30" (76 cm) Appx 5.5 mL 20 Drop Admin Set w/Integrated Clave Drip Chamber, 1.2 Micron Filter, Spiros, Bag Hanger, REF:
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 3mL Syringe w/Spinning Spiros, Red Cap, REF: CH2003; 5mL Syringe w/Spinning Spiros, Red Cap, REF: CH2005; 10mL Syringe w
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger, REF: CL3011; 3
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integ
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls Spinning Spiros, Closed Male Luer, REF: 060-CH2000S; SURPLUG ChemoAccess Closed Male Luer, REF: 066-CH2000S; SURPLUG Che
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 30, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave D
Hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
December 9, 2020Medical DevicesFDA
ICU Medical, Inc. recalls REF MC330209 6' (15 cm) Appx 0.09 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear, 0.2 Micron Filter, 1.2 Micron Filt
Hazard: Incorrect filter assemblies in IV administration set.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls SURPLUG Nano Connector.1 unit per pouch, 100 multivac pouches per case. UDI:(01)54987892069913(17)221201(30)100(10)44764
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 10 IN (25cm) APPX 1.1ml, SMALLBORE EXT w/3-Port NanoClave Manifold, Check Valve. 1 unit per pouch, 50 pouches per case.
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 5.5" (14 cm) Appx 0.28 ml, Smallbore Ext Set w/NanoClave, Clamp, Luer Lock. 1 unit per pouch, 50 multivac pouches per ca
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 7 IN(18 cm)APPX 1.2ml. EXT SET, NanoClave T-CONNECTOR, CLAMP, ROTATING LUER. 1 unit per pouch, 50 pouches per case. UDI:
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 8" (20 cm) Appx 0.33 ml, Smallbore Ext Set w/MicroClave Clear, NanoClave Clamp, Rotating Luer.1 unit per pouch, 50 multi
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 226 cm (89 IN) APPX 2.7ml, MicroClave Clear, 1.2v FILTER Nanoclave T-CONN. 1 unit per pouch, 50 pouches per case. UDI:
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 58cm (23 IN) APPX 2.1 ml EXT SET Smallbore PUR Yellow w/6-Port Clave, Clave, 2 Clave Stopcock. 1 unit per pouch, 50 pouc
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls PROXIMALE, 28 cm (11 IN) APPX 1.0ml, EXT Set, 6-PORT NanoClave MANIFOLD, NanoClave. 1 unit per pouch, 50 pouches per cas
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls APPX 0.75ml, 6-Port NanoClave Manifold (Glow, Red, Blue, Purple, Yellow Rings)BCV. 1 unit per pouch, 50 pouches per case
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls DISTALE - 203 cm (80") Bifuse Ext Set w/6 Gang 1o2 Manifold, 8 MicroClave Clear (Green Rings), Back Check Valve, 3-Port
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 165cm (65 IN)APPX 2.2ml, PUR EXT SET, 3-PORT NanoClave MANIFOLD, BCV, NanoClave. 1 unit per pouch, 50 pouches per case.
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 20 cm (8 IN) APPX 1.0ml, PUR EXT SET, 3-PORT, NanoClave MANIFOLD, BCV, NanoClave.1 unit per pouch, 50 pouches per case.
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 15 cm (6 IN) Appx 1.1 ml, Smallbore Quadfuse Ext Set w/4 NanoClave 3 BCV RLL.1 unit per pouch, 50 pouches per case. UD
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 135 cm (53") Ext Set w/Remv 5 Gang Stopcock, Baseplate w/7 NanoClave, 4-Way Stopcock, Rotating Luer, 1 Ext..1 unit per p
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 8 IN(20cm) APPX 0.46ml, MicroClave CLEAR, NanoClave T-CONN, FILTER. 1 unit per pouch, 50 pouches per case. UDI: (01)108
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
October 28, 2020Medical DevicesFDA
ICU Medical, Inc. recalls 7" (18 cm) Appx 0.29 ml, Smallbore Ext Set w/MicroClave Clear, NanoClave, Clamp, Rotating Luer. 1 unit per pouch, 50 mul
Hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.