ImaCor Recalls
2 recalls on record · November 6, 2024 to October 15, 2025 · Medical Devices
Recalls by year
24
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
October 15, 2025Medical DevicesFDA
ImaCor Inc. recalls ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultra
Hazard: Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
November 6, 2024Medical DevicesFDA
ImaCor Inc. recalls ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Mo
Hazard: Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.