RecallWatchUSA

Home / Brands / In2Bones, SAS

In2Bones, SAS Recalls

10 recalls on record · April 1, 2015 to March 27, 2024 · Medical Devices

Recalls by year

15
17
21
24

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

March 27, 2024Medical DevicesFDA
In2Bones, SAS recalls I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 80mm,Part Number S65 ST180
Hazard: This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.
March 27, 2024Medical DevicesFDA
In2Bones, SAS recalls I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 75mm,Part Number S65 ST175
Hazard: This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.
March 31, 2021Medical DevicesFDA
In2Bones, SAS recalls In2Bones Shannon Percutaneous Burr - Product Usage: indicated for forefoot mini-invasive and percutaneous surgery.
Hazard: This Field Action is being conducted following identification of manufacturing non-conformity of some batches of Shannon burrs diameter 2.0mm length 20mm, possibly causing the breakage of the burr during use.
August 9, 2017Medical DevicesFDA
In2Bones, SAS recalls Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 S
Hazard: The firm received several customer complaints reporting some perioperative breakages of NE0411 Non-Locking screws when inserting the screw into the radius diaphysis
January 18, 2017Medical DevicesFDA
In2Bones, SAS recalls NEOVIEW(R) SCREWS SET, REF W27 ST900, STERILE R The NEOVIEW¿ Plating System is intended for fixation of intra-articula
Hazard: Dimensional non-conformity of the thread on the screw head.
January 18, 2017Medical DevicesFDA
In2Bones, SAS recalls NEO SCREW 02.7-14MM LOCKING, REF W27 ST114, In2Bones, 28 chemin du petit bois, 69130 Ecully - France The NEOVIEW¿ Plat
Hazard: Dimensional non-conformity of the thread on the screw head.
January 18, 2017Medical DevicesFDA
In2Bones, SAS recalls SCREW 02.7-12MM LOCKING, REF W27 ST112: a) DRP, b) NEO, In2Bones, 28 chemin du petit bois, 69130 Ecully - France The
Hazard: Dimensional non-conformity of the thread on the screw head.
December 16, 2015Medical DevicesFDA
In2Bones, SAS recalls IBS - Osteosynthesis screws for bone fixation Model S25 ST014.
Hazard: One batch of osteosynthesis screws were manufactured with the incorrect Torx imprint on the screw head.
April 1, 2015Medical DevicesFDA
In2Bones, SAS recalls DUAFIT(R) PIP 0o CA SIZE 3, REF A60 SP003, STERILE R, Rx only. Interphalangeal implant.
Hazard: Dimensional non-conformity leading to the impossibility to introduce the implant in the dedicated holder.
April 1, 2015Medical DevicesFDA
In2Bones, SAS recalls DUAFIT(R) PIP 0o CA SIZE 2, REF A60 SP002, STERILE R, Rx only. Interphalangeal implant.
Hazard: Dimensional non-conformity leading to the impossibility to introduce the implant in the dedicated holder.
In2Bones, SAS Recalls | RecallWatchUSA