RecallWatchUSA

Home / Brands / Integra

Integra Recalls

9 recalls on record · July 25, 2012 to May 22, 2019 · Medical Devices

Recalls by year

12
13
19

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

May 22, 2019Medical DevicesFDA
Integra Limited recalls Integra Flowable Wound Matrix Catalog No: FWD301 Product Usage: Integra Flowable Wound Matrix is indicated for the
Hazard: Stability did not meet acceptance criteria for visual appearance during mixing and force vs. displacement at 6 months
April 17, 2019Medical DevicesFDA
Integra Limited recalls Integra bioBLOCK Resorbable subtaler Implant, 12mm Catalog Number: 040012 - Product Usage: The bioBLOCK Implant is in
Hazard: Temperature indicator may have changed color to dark gray or black prior to shipment. Change in color may indicate the product exposure to temperatures greater than 110¿F
April 17, 2019Medical DevicesFDA
Integra Limited recalls Integra bioBLOCK Resorbable subtaler Implant, 11mm Catalog Number: 040011 - Product Usage: The bioBLOCK Implant is in
Hazard: Temperature indicator may have changed color to dark gray or black prior to shipment. Change in color may indicate the product exposure to temperatures greater than 110¿F
April 17, 2019Medical DevicesFDA
Integra Limited recalls Integra bioBLOCK Resorbable subtaler Implant, 10mm Catalog Number: 40010 - Product Usage: The bioBLOCK Implant is ind
Hazard: Temperature indicator may have changed color to dark gray or black prior to shipment. Change in color may indicate the product exposure to temperatures greater than 110¿F
November 6, 2013Medical DevicesFDA
Integra Limited recalls Brand Names: Integra¿ Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B,
Hazard: Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA requirements, certain Camino¿ Intracranial Pressure Monitoring Kits require additional labelling to indicate they are UNSAFE in an MR environment. Only the following Camino¿ Intracranial Pressure Monitoring Kits are affected: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.
September 4, 2013Medical DevicesFDA
Integra Limited recalls Integra Leyla Ball Joint Clamp, also as included in Leyla Table Mounting Hardware. The Leyla Retraction System, as p
Hazard: As a result of complaint investigations it was identified that there is a potential that the Leyla Ball Joint Clamp (R2383) may not provide the level of stability needed to maintain retraction in all cases.
October 3, 2012Medical DevicesFDA
Integra Limited recalls The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for
Hazard: An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a report of patient injuries or an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that there is no potential harm to patients, onl
October 3, 2012Medical DevicesFDA
Integra Limited recalls The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for
Hazard: An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a report of patient injuries or an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that there is no potential harm to patients, onl
July 25, 2012Medical DevicesFDA
Integra Limited recalls Integra(TM) Jarit(R) Take-Apart Endoscopic Instruments. Integra(TM) Jarit(R) Take-Apart Endoscopic Instruments are in
Hazard: The Integra(TM) Jarit(R) Take-Apart Laparoscopic Instrument shafts were released with internal component out of specification.
Integra Recalls | RecallWatchUSA