Home / Brands / INTELERAD MEDICAL SYSTEMS INCORPORATED
INTELERAD MEDICAL SYSTEMS INCORPORATED Recalls
2 recalls on record · March 5, 2025 to December 3, 2025 · Medical Devices
Also filed under
The recall feeds record what looks like the same company under more than one name:
December 3, 2025Medical DevicesFDA
INTELERAD MEDICAL SYSTEMS INCORPORATED recalls IntelePACS - InteleConnect / TechPortal
Hazard: Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.
March 5, 2025Medical DevicesFDA
INTELERAD MEDICAL SYSTEMS INCORPORATED recalls IntelePACS (Image Fusion Module) - InteleViewer
Hazard: A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.