July 29, 2026Medical DevicesFDA
KENT IMAGING, INC. recalls SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
Hazard: The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
January 31, 2024Medical DevicesFDA
KENT IMAGING, INC. recalls Snapshot NIR, REF: KD204
Hazard: Tissue oxygenation measurement system touchscreens may be unresponsive, which may prevent the viewing of oxygenation images.