Lab Vision Recalls
3 recalls on record · January 22, 2014 to February 24, 2016 · Medical Devices
Recalls by year
14
15
16
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
February 24, 2016Medical DevicesFDA
Lab Vision Corporation recalls CEA / CD66e Ab-2, Product code RB-368-R7, RB-368-R1; 7 ml each. Description Carcinoembryonic antigen (CEA), is sy
Hazard: Certain lots of CEA/CD66e Ab2 were produced with a dilution factor of 1:100 instead of the recommended 1:250.
December 2, 2015Medical DevicesFDA
Lab Vision Corporation recalls Thermo Scientific p21 WF1 Ab-3 (DCS-60.2) 1 ml (0.4mg/ml): Product code: MS-230-P and MS-230P0; Hematology: p21WAF1
Hazard: It was discovered that the labeling of MS-230 antibody incorrectly describes the concentration - label reads 0.2mg/ml but it is actually 0.4 mg/ml.
January 22, 2014Medical DevicesFDA
Lab Vision Corporation recalls Estrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision Corporation,
Hazard: Reports that the product is not performing as expected. In house testing showed no staining on breast cancer tissues.