Home / Brands / Laborie Medical Technologies
Laborie Medical Technologies Recalls
2 recalls on record · July 31, 2019 to July 8, 2026 · Medical Devices
Recalls by year
19
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
July 8, 2026Medical DevicesFDA
Laborie Medical Technologies Corp recalls Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar
Hazard: Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.
July 31, 2019Medical DevicesFDA
LABORIE MEDICAL TECHNOLOGIES, CORP recalls U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract d
Hazard: Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard