Home / Brands / Laserex Systems
Laserex Systems Recalls
2 recalls on record · April 13, 2016 to June 14, 2017 · Medical Devices
Recalls by year
16
17
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
June 14, 2017Medical DevicesFDA
Laserex Systems Inc. recalls Ellex Tango Reflex with slit lamp Laser Ophthalmic In the SLT mode, the device is intended to be used for selective
Hazard: It was discovered the unit produced a laser emission without pressing the fire button when the slit-lamp was driven to its lowest position prior to use on any patient.
April 13, 2016Medical DevicesFDA
Laserex Systems Inc. recalls Ellex Medical Fixed eye Safety filters Model Integre S LP5532. The filter is located in the binocular mount of the l
Hazard: The Firm discovered that there is a design error which is traceable back to 2012 when a design change on the safety filter optic adaptor was implemented.