Home / Brands / LEONI CIA CABLE SYSTEMS
LEONI CIA CABLE SYSTEMS Recalls
2 recalls on record · November 29, 2023 to November 29, 2023 · Medical Devices
November 29, 2023Medical DevicesFDA
LEONI CIA CABLE SYSTEMS recalls ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation
Hazard: Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.
November 29, 2023Medical DevicesFDA
LEONI CIA CABLE SYSTEMS recalls ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation
Hazard: Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.