Home / Brands / Linet Spol. S.r.o
Linet Spol. S.r.o Recalls
3 recalls on record · March 20, 2019 to March 15, 2023 · Medical Devices
Recalls by year
19
21
23
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
March 15, 2023Medical DevicesFDA
Linet Spol. S.r.o. recalls Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1.
Hazard: Labeling being updated to assure user is securing the siderail to a locked position to prevent potential patient fall and injury.
January 13, 2021Medical DevicesFDA
Linet Spol. S.r.o. recalls Eleganza 5 (AC powered adjustable bed), Item Numbers: a) 1GE512055-40, b) 1GE512055-45, c) 1GE512055-85, and d) 1GE5120
Hazard: A component used within the side rail mechanism may malfunction, resulting in the side rail either not locking properly in the upper position or the side rail becoming stuck in the upper position.
March 20, 2019Medical DevicesFDA
Linet Spol. S.r.o. recalls AVE 2 Birthing Bed
Hazard: LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the orange insulation of the power cable may fail causing a possible spark or possible exposure to a live conductor.