Home / Brands / Mechatronics USA
Mechatronics USA Recalls
2 recalls on record · June 14, 2023 to June 14, 2023 · Medical Devices
June 14, 2023Medical DevicesFDA
Mechatronics USA recalls RR Mechatronics Control Level A ESR Control4.6 ml QRR049002- In-vitro diagnostic quality control material to monitor t
Hazard: QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.
June 14, 2023Medical DevicesFDA
Mechatronics USA recalls RR Mechatronics Starrsed Control Level N ESR Control 4.6 ml QRR049001- In-vitro diagnostic quality control material to m
Hazard: QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.