February 25, 2026Medical DevicesFDA
Medica Corporation recalls Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
Hazard: A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
May 15, 2019Medical DevicesFDA
Medica Corporation recalls Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Us
Hazard: Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
April 4, 2018Medical DevicesFDA
Medica Corporation recalls EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for e
Hazard: Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.
March 28, 2018Medical DevicesFDA
Medica Corporation recalls Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , RE
Hazard: The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.