Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Recalls

198 recalls on record · February 28, 2018 to June 28, 2023 · Medical Devices

Recalls by year

18
19
20
21
22
23

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic EnTrust Escudo D144VRC (OUS distribution) Product Usage: The implantable cardioverter defibrillator is inte
Hazard: Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRe
Hazard: Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRe
Hazard: Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic EnTrust Dual Chamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRea
Hazard: Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic EnTrust Dual Chamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRea
Hazard: Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Percepta Quad CRT-P MRI SureScan, REF W4TR04 Product Usage: The CRT-P models are implantable pulse generators with c
Hazard: Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Percepta CRT-P MRI SureScan, REF W1TR04 Product Usage: The CRT-P models are implantable pulse generators with cardia
Hazard: Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Percepta Quad CRT-P MRI SureScan, REF W4TR01 Product Usage: The CRT-P models are implantable pulse generators with c
Hazard: Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
July 18, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Percepta CRT-P MRI SureScan, REF W1TR01 Product Usage: The CRT-P models are implantable pulse generators with cardia
Hazard: Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
July 4, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Single Chamber Temporary External Pacemaker, Model 53401
Hazard: A subset of Medtronic Model 53401 Single Chamber External Pulse Generators (EPGs) may revert from programmed settings to nominal settings during use with a patient due to an unexpected Power on Reset (POR).
February 28, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Implantable Cardioverter Defibrillators (ICDs), Visia AF. Labeled as: a. Visia AF VR ICD DF1 (Product No. DVAB1D1);
Hazard: Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
February 28, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Implantable Cardioverter Defibrillators (ICDs), Visia AF MRI. Labeled as the following: a. Visia AF MRI VR SureScan IC
Hazard: Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
February 28, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Implantable Cardioverter Defibrillators (ICDs), Evera. Labeled as the following: a. Evera S DR ICD DF1 (Product No. DD
Hazard: Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
February 28, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Claria MRI. Labeled as the following: a. Claria MRI CRT-D
Hazard: Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
February 28, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Implantable Cardioverter Defibrillators (ICDs), Evera MRI. Labeled as the following: a. Evera MRI XT DR ICD DF1 (Pro
Hazard: Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
February 28, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Viva. Labeled as the following: a. Viva Quad S CRT-D DF4
Hazard: Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
February 28, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Compia MRI. Labeled as the following: a. Compia MRI CRT-D
Hazard: Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
February 28, 2018Medical DevicesFDA
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Amplia MRI. Labeled as the following: a. Amplia MRI CRT-D
Hazard: Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).