Home / Brands / Mercury Enterprises, Inc.
Mercury Enterprises, Inc. Recalls
3 recalls on record · January 20, 2016 to January 22, 2025 · Medical Devices
Recalls by year
16
24
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
January 22, 2025Medical DevicesFDA
Mercury Enterprises, Inc. dba Mercury Medical recalls Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 105
Hazard: Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
October 9, 2024Medical DevicesFDA
Mercury Enterprises, Inc. dba Mercury Medical recalls Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.
Hazard: Manual resuscitator circuit flow controller may come apart due to a missing manufacturing step.
January 20, 2016Medical DevicesFDA
Mercury Enterprises, Inc. dba Mercury Medical recalls Mercury Medical Neo-Tee T-Piece Resuscitator and T-Piece Circuit. The Neo-Tee T-Piece Resuscitator is a gas powered eme
Hazard: Incorrect use of T-Piece resuscitators will result in complete occlusion of the patient circuit and consequent failure to ventilate the patient.