Moller Vital Recalls
5 recalls on record · May 29, 2013 to May 22, 2019 · Medical Devices
Recalls by year
13
16
19
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
May 22, 2019Medical DevicesFDA
Moller Vital recalls Molift 2-Point sling bar, an accessory for Molift Mover 180/Air. Item nos. 1830001 1830002 1830002C 1830003
Hazard: The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.
May 22, 2019Medical DevicesFDA
Moller Vital recalls Molift 2-Point sling bar, an accessory for Molift Mover 205. Item nos. 1430016 1430017 1430021
Hazard: The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.
August 3, 2016Medical DevicesFDA
Moller Vital recalls Molift Air 205, Molift Air 205 Propulsion, Molift Air IRC, Molift Air 205 IRC Propulsion, Molift Air 300, Molift Air 300
Hazard: Potential for the bolt on the suspension coupling on the Molift Air to come loose from the housing.
May 29, 2013Medical DevicesFDA
Moller Vital recalls Molift Smart 150
Hazard: The leg spreading mechanism will fail to hoist and collapse if assemble incorrectly by the user.
May 29, 2013Medical DevicesFDA
Moller Vital recalls Molift Smart 150
Hazard: The leg spreading mechanism will fail to hoist and collapse if assemble incorrectly by the user.