1 recall on record · August 24, 2022 to August 24, 2022 · Medical Devices
August 24, 2022
Medical Devices
FDA
Myolyn Inc. recalls Power electronic box, Part #AS-0066, subassembly for the MyoCycle MC-2.
Hazard: Screws mounting an internal power supply may come loose during shipping resulting in increased risk of high voltage electrical shock.