NovaSignal Recalls
4 recalls on record · February 8, 2023 to July 3, 2024 · Medical Devices
Recalls by year
23
24
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
July 3, 2024Medical DevicesFDA
NovaSignal Corp. recalls NovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2
Hazard: Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.
July 3, 2024Medical DevicesFDA
NovaSignal Corp. recalls NovaGuide 2 Intelligent Ultrasound, REFs: NSC-NVGSYS2 & NSC-NVGSYS2-CA
Hazard: Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.
February 8, 2023Medical DevicesFDA
NovaSignal Corp. recalls NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1. Comprised of NovaBot Automated Headset (AKA: NeuralBot, Mode
Hazard: Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.
February 8, 2023Medical DevicesFDA
NovaSignal Corp. recalls Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System
Hazard: Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.