Ohio Medical Recalls
4 recalls on record · July 18, 2018 to June 25, 2025 · Medical Devices
Recalls by year
18
22
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
June 25, 2025Medical DevicesFDA
Ohio Medical Corporation recalls GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S
Hazard: Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.
January 19, 2022Medical DevicesFDA
Ohio Medical Corporation recalls Instavac Portable Suction Pump, Constant mode (model #757000, Instavac C) and Intermittent mode (model 756000, Instavac
Hazard: Increase in premature device failures
September 19, 2018Medical DevicesFDA
Ohio Medical Corporation recalls Portable Suction Pump, (a) Style 758010, Model CARE-E-VAC 3, Ohio Medical LLC; (b) Style 758000, Model CARE-E-VAC 3, Oh
Hazard: The care-e-vac(R)3 Battery/AC Powered Portable Aspirator battery terminal may come in contact with the pump causing an electric short leading to damage to wiring and resulting in loss of function or an inoperable unit under battery power.
July 18, 2018Medical DevicesFDA
Ohio Medical Corporation recalls Vacuum Regulator, Push-T-Set Intermittent Suction Unit (PTS-ISU) Product Usage: The vacuum regulator is intended to
Hazard: Final quality control testing was not completed before the devices were distributed..