RecallWatchUSA

Home / Brands / Outset Medical

Outset Medical Recalls

5 recalls on record · March 30, 2022 to October 9, 2024 · Medical Devices

Recalls by year

22
23
24

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

October 9, 2024Medical DevicesFDA
Outset Medical, Inc. recalls Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indi
Hazard: Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.
June 26, 2024Medical DevicesFDA
Outset Medical, Inc. recalls TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747
Hazard: A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.
April 24, 2024Medical DevicesFDA
Outset Medical, Inc. recalls Tablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001
Hazard: Hemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may contain non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs), that have not been flushed out over time with usage. Exposure to PCBAs could cause: skin conditions, liver damage, impaired reproduction, development of certain kinds of cancer in humans.
April 19, 2023Medical DevicesFDA
Outset Medical, Inc. recalls Outset Tablo Model Numbers (REF): PN-0006000 PN-0007001 Tablo is a next generation, self-contained hemodialysis
Hazard: An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.
March 30, 2022Medical DevicesFDA
Outset Medical, Inc. recalls Outset Tablo Console Model Number: PN-0003000 Model Number: PN-0006000
Hazard: Due to a component in the hemodialysis console there is the possibility of heat-related damage that may occur with the device.