December 26, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Xper Flex Cardio Physiomonitoring System, Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 4535
Hazard: There are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.
December 19, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 - Product Usage: Indicated for use by health
Hazard: A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
October 24, 2018Medical DevicesFDA
Philips Electronics North America Corp. recalls Philips HeartStart FR3 Defibrillator, Model: 861388, 861389 Product Usage: The HeartStart FR3 is used to treat suspe
Hazard: Automated external defibrillators may not fully meet IPx5 water ingress specification. The device may fail to function should water intrusion occur.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Cesar Powerpack-Visub(V3000), System Code 72243 Vascular, cardiovascular and neurovascular imaging applications, Car
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Cesar-DMCP-Visub(HM2000/3000), System Code 72239 Vascular, cardiovascular and neurovascular imaging applications, C
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Poly C- DMCP-Visub(H3000), System Code 72238 Vascular, cardiovascular and neurovascular imaging applications, Cardia
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper F020 Biplane DR Table, System Code 722020 Vascular, cardiovascular and neurovascular imaging application
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper F010/10 DRTable, System Code 722019 Vascular, cardiovascular and neurovascular imaging applications, Car
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper F020 DR Table, System Code 722015 Vascular, cardiovascular and neurovascular imaging applications, Cardi
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper F010 DR Table, System Code 722014 Vascular, cardiovascular and neurovascular imaging applications, Cardi
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper F020 DR Table, System Code 722023 Vascular, cardiovascular and neurovascular imaging applications, Cardi
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls AIIura Xper F010 DR Table, System Code 722022 Vascular, cardiovascular and neurovascular imaging applications, Cardi
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls AIIuraXperFDlO/10, System Code 722011 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imagi
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls AIIuraXperFDlO, System Code 722010 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls AlluraXperFOlO/lO, System Code 722005 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imagi
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper F010, System Code 722003 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imagin
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls AlIura Xper F010 F, System Code 722002 Vascular, cardiovascular and neurovascular imaging applications, Cardiac ima
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper FD1O C, System Code 722001 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imag
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper FD2O Biplane, System Code 722008 Vascular, cardiovascular and neurovascular imaging applications, Cardia
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper FD2O Biplane, System Code 722013 Vascular, cardiovascular and neurovascular imaging applications, Cardia
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper FD2O, System Code 722012 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imagin
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls Allura Xper FD2O, System Code 722006 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imagin
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls NTEGRIS Allura 9 0 FDXD, System Code 722498 Vascular, cardiovascular and neurovascular imaging applications, Cardia
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls INTEGRIS Allura 9 F FDXD, System Code 722497 Vascular, cardiovascular and neurovascular imaging applications, Cardia
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls INTEGRIS SUITE, System Code 722199 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls INTEGRIS Allura 15-12 (biplane), System Code 722044 Vascular, cardiovascular and neurovascular imaging applications,
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls INTEGRIS Allura 15-12 (mono), System Code 722043 Vascular, cardiovascular and neurovascular imaging applications, Ca
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls INTEGRIS CV, System Code 722030 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging app
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls INTEGRIS Allura 9 (biplane), System Code 722021 Vascular, cardiovascular and neurovascular imaging applications, Car
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
October 3, 2018Medical DevicesFDA
Philips Electronics North America Corporation recalls INTEGRIS Allura 9, System Code 722018 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imagi
Hazard: There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.