Pointe Scientific Recalls
3 recalls on record · September 9, 2015 to September 23, 2015 · Medical Devices
September 23, 2015Medical DevicesFDA
Pointe Scientific, Inc. recalls Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and
Hazard: Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will be out of specifications.
September 23, 2015Medical DevicesFDA
Pointe Scientific, Inc. recalls Pointe Scientific autoHDL Reagent H7545 H7545-R1 in a bulk cube container or HDPE plastic bottle H7545-1000 kit incl
Hazard: Crystals may be observed in the R1 reagent. Crystals may be a result of the reagent being frozen during transport. The crystals do not impact product performance.
September 9, 2015Medical DevicesFDA
Pointe Scientific, Inc. recalls Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also so
Hazard: A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.