December 7, 2022Medical DevicesFDA
Primus Corporation dba Trinity Biotech recalls Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis
Hazard: The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. Additionally, per the firm, the unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.
December 7, 2022Medical DevicesFDA
Primus Corporation dba Trinity Biotech recalls A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052,
Hazard: The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.