RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) Recalls

50 recalls on record · July 8, 2015 to March 17, 2021 · Medical Devices

Recalls by year

15
16
20
21

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 30mm Material Number: 66-230 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 1 Level, 22mm Material Number: 66-122 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 5 Level, 95mm Material Number: 66-595 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 5 Level, 90mm Material Number: 66-590 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 5 Level, 85mm Material Number: 66-585 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 5 Level, 80mm Material Number: 66-580 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 5 Level, 75mm Material Number: 66-575 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 5 Level, 105mm Material Number: 66-5105 - Product Usage: intended fo
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 5 Level, 100mm Material Number: 66-5100 - Product Usage: intended fo
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 4Level, 88mm Material Number: 66-488 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 4Level, 84mm Material Number: 66-484 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 4Level, 80mm Material Number: 66-480 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 4Level, 76mm Material Number: 66-476 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 4Level, 72mm Material Number: 66-472 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 4Level, 68mm Material Number: 66-468 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 4Level, 64mm Material Number: 66-464 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 4Level, 60mm Material Number: 66-460 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 3Level, 63mm Material Number: 66-363 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 3Level, 60mm Material Number: 66-360 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 3Level, 57mm Material Number: 66-357 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 3Level, 54mm Material Number: 66-354 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 3Level, 51mm Material Number: 66-351 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 3 Level, 48mm Material Number: 66-348 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 3 Level, 45mm Material Number: 66-345 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 3 Level, 42mm Material Number: 66-342 - Product Usage: intended for
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 42mm Material Number: 66-242 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 40mm Material Number: 66-240 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 38mm Material Number: 66-238 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 36mm Material Number: 66-236 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
March 17, 2021Medical DevicesFDA
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) recalls CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 34mm Material Number: 66-234 - Product Usage: intended for a
Hazard: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision