March 23, 2022Medical DevicesFDA
Sam Medical Products recalls SAM CHEST SEAL REF CS203-EN Combo LOT Y060321-09
Hazard: Due to partially assembled valved dressing that has a hole within it instead of the intended non-valved dressing in the package resulting in that the dressing cannot be centered over the open chest wound and create an occlusive seal due to the hole being present.
November 5, 2014Medical DevicesFDA
Sam Medical Products recalls SAM Junctional Tourniquet (SJT) Auxiliary, Model SJT 102 and SJT 112. The SJT models specific intended uses are to c
Hazard: SAM Junctional Tourniquet (SJT) Auxiliary strap, model SJT 102 and SJT 112, is recalled because the snap hook on auxiliary strap broke during training demonstration.