Signal Medical Recalls
3 recalls on record · October 29, 2014 to April 26, 2017 · Medical Devices
Recalls by year
14
17
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
April 26, 2017Medical DevicesFDA
Signal Medical Corporation recalls Soft Face "No Bounce" Mallet Model/Catalog: BIO-1-1090
Hazard: Potential for poly tip of the mallet to come off during use. Potential for foreign material in the surgical field or delay in surgery. Reminder for customers that the poly tip must be removed for cleaning and sterilization to assure proper cleaning and sterilization of the entire mallet.
October 29, 2014Medical DevicesFDA
Signal Medical Corporation recalls Symmetric Total Knee Femoral Components Product Usage: Symmetric Total Knee Femoral Component is intended for use wi
Hazard: Packaging of a device was compromised
October 29, 2014Medical DevicesFDA
Signal Medical Corporation recalls Microseal Augmented (20 degrees) Liner, acetabular liner component made from UHMW polyethylene (ASTM F648). Sizes 36x52x
Hazard: Incorrect dimension on four liners was detected