Home / Brands / Sorin Group USA
Sorin Group USA Recalls
52 recalls on record · October 31, 2012 to November 15, 2017 · Medical Devices
Recalls by year
12
13
15
16
17
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
November 15, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls FlexFlow Venous Cannula Model/Catalog Number 200-200 The FlexFlow Venous Cannula is a single use sterile device. It is
Hazard: A specific lot of the FlexFlowTM Venous Cannula, 200-200, is being recalled because it is sharper or pointier than expected.
August 16, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls Sorin Group Aortic Arch Cannula, 7 mm x 10 in x 3/8 in, Rx Only, Sterile Product Usage: The Aortic Arch Cannulae are
Hazard: Identification of excess plastic on the tip of the cannula.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls NON-STERILE SAMPLE, SORIN GROUP, SMART PERFUSION PACK, REF 627376401, 1 EA, NOT FOR CLINICAL USE
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls NON-STERILE SAMPLE, SORIN GROUP, SMART CARDIOPLEGIA, CARDIOPLEGIA PACK, REF 627366901, 1 EA, NOT FOR CLINICAL USE
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls NON-STERILE SAMPLE, SORIN GROUP, SMART PERFUSION PACK, INSPIRE 8FS, REF 044043700, 1 EA, NOT FOR CLINICAL USE
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, SMART 1/4" PEDS PACK, REF 627262502, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, INSPIRE 8S, REF 044036101, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, TOTAL VISION SYSTEM-SMA, REF 084510101, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, NEONATE/PEDIATRIC PACK, REF 627127702, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls STERILE SAMPLE SORIN GROUP, SMART PERFUSION PACK, NEONATE/PEDIATRIC PACK, REF 627127703, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, SMART 3/16" PEDS PACK, REF 627263601, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, PUMP AND TABLE PACK, REF 627374101, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, INSPIRE 8FS S5 PACK W/CARDIOPLEGIA-3D, (a) REF 044042600 AND (b) REF 044011200, STERI
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, SMART VENOUS VACUUM, REF 084118102, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, INSPIRE 6FS, REF 044008204, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, SMA PRIMO2X OPTI PACK, REF 088512900, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, SMART CIRCULATORY ASSIST PACK, REF 627255003, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 9, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls SORIN GROUP, SMART PERFUSION PACK, LONG PACK, REF 088512600, STERILE EO, Rx Only, 1 EA
Hazard: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
August 2, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls STOCKERT Heater-Cooler System 3T, 200V/50 Hz/60 Hz
Hazard: Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
August 2, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls STOCKERT Heater-Cooler System 3T, 120V/60 Hz
Hazard: Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
August 2, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls STOCKERT Heater-Cooler System 3T, 127V/60 Hz
Hazard: Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
August 2, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls STOCKERT Heater-Cooler System 3T, 208V/60 Hz
Hazard: Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
August 2, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls STOCKERT Heater-Cooler System 3T, 240V/60 Hz
Hazard: Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
August 2, 2017Medical DevicesFDA
Sorin Group USA, Inc. recalls STOCKERT Heater-Cooler System 3T, 230 V
Hazard: Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
December 14, 2016Medical DevicesFDA
Sorin Group USA, Inc. recalls Perceval Sutureless Aortic Heart Valve size S, Item # ICV1208, REF PVS21; Perceval Sutureless Aortic Heart Valve size M,
Hazard: Sorin Group USA, Inc. announces a voluntary field action for the Perceval Sutureless Heart Valve due to updated steps associated with the implantation procedure.
December 7, 2016Medical DevicesFDA
Sorin Group USA, Inc. recalls ST¿CKERT HeaterCooler System 3T; Item Number 160281, HeaterCooler 3T, 240V/60Hz; Item Number 160282, HeaterCooler
Hazard: On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).
November 23, 2016Medical DevicesFDA
Sorin Group USA, Inc. recalls The ORCHESTRA PLUS Programmer A portable device, equipped with a microprocessor used to program and interrogate Sori
Hazard: Sorin Group USA, Inc. announces a voluntary field action for the ORCHESTRA Programmer and ORCHESTRA PLUS Programmer because the residual longevity estimate may be less accurate.
November 23, 2016Medical DevicesFDA
Sorin Group USA, Inc. recalls The ORCHESTRA Programmer A portable device, equipped with a microprocessor used to program and interrogate Sorin imp
Hazard: Sorin Group USA, Inc. announces a voluntary field action for the ORCHESTRA Programmer and ORCHESTRA PLUS Programmer because the residual longevity estimate may be less accurate.
September 28, 2016Medical DevicesFDA
Sorin Group USA, Inc. recalls St¿ckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The St¿c
Hazard: Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.
September 14, 2016Medical DevicesFDA
Sorin Group USA, Inc. recalls Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloo
Hazard: The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.