Home / Brands / Stihler Electronic
Stihler Electronic Recalls
3 recalls on record · January 30, 2019 to January 30, 2019 · Medical Devices
January 30, 2019Medical DevicesFDA
Stihler Electronic Gmbh recalls Heating Profile PRISMAFLO II blood return warmer for Prismaflex Product Code: PF2-WP2618 Product Usage: The PR
Hazard: PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
January 30, 2019Medical DevicesFDA
Stihler Electronic Gmbh recalls Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP31 Product Usage: The PRI
Hazard: PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
January 30, 2019Medical DevicesFDA
Stihler Electronic Gmbh recalls Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP33 Product Usage: The PRISM
Hazard: PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn