RecallWatchUSA

Home / Brands / Synthes (USA) Products

Synthes (USA) Products Recalls

164 recalls on record · August 1, 2012 to April 8, 2026 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
22
23
24
25
26

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

June 15, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls 1.5 mm Rapid Resorb Orbt1 F1 P1 w/Bending Template / 24 mm - Sterile, Rapid Resorbable Fixation System. 1.5 mm Rapid
Hazard: Labeling errors. The outer packge label expiration date may exceed the expiration date of the subcomponents.
June 15, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43
Hazard: Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points. This could potentially be reproduced during use and reprocessing.
June 1, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls 3.2 mm three-fluted Radiolucent Drill Bit/Needle Point/145 mm Sterile Product Usage: The 3.2 mm Three-fluted Drill B
Hazard: The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.
June 1, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls 3.2 mm Three-fluted drill bit, sterile, Catalog ID # 03.010.060S Product Usage: Usage: The 3.2 mm Three-fluted Dril
Hazard: The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.
May 4, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls 5.0 mm Periprosthetic Locking Screw Self-tapping 12 mm Sterile; Catalog ID # 02.221.462S Indicated for fixation of var
Hazard: This lot of the 5.0 mm Periprosthetic Locking Screw, Self-tapping, 12mm Sterile, was shipped after the expiration date of May 31, 2012. The label on the product has the correct expiration date.
April 27, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls Cervical Spine Locking Plate (CSLP) System, 4mm Cancellous Self Tapping Screw, Length 16mm Lot # 9666776 and 4mm Cortex
Hazard: DePuy Synthes is initiating a Voluntary Medical Device Recall of lot # 9666776 and 9671879 of Cancellous and Cortex 4.0 mm Screws, which are a part of the Cervical Spine Locking Plate (CSLP) System due to discovery of a single lot of 450.136 screws (4mm Cancellous Self Tapping Screw, Length 16mm, lot 9666776) being swapped with a single lot of 450.016 (4mm Cortex Standard Screw, length 16 mm, lot 9671879).
April 20, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls Oblique Posterior Atraumatic Lumbar (OPAL) System, Polymeric spinal fusion cage, non-sterile, OPAL Spacer 10 mm x 24 mm,
Hazard: The label on the outer packaging does not match the product inside and intended use of the device.
April 20, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D
Hazard: It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.
April 20, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D
Hazard: It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.
April 20, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D
Hazard: It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.
April 20, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D
Hazard: It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.
April 20, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D
Hazard: It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.
April 13, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile
Hazard: Out of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.
April 13, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile
Hazard: Out of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.
April 6, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls Synthes Interlocking Bolt, 151 mm (General and Plastics Surgery Devices) Guide, Surgical Instrument
Hazard: DePuy Synthes is initiating a voluntary medical device recall of certain lots of the Interlocking Bolt which are a part of the Synthes 4.5 mm LCP Proximal Tibia Plate System. The Synthes 4.5 mm LCP Proximal Tibia Plate are indicated for the treatment of nonunions, malunions and fractures of the proximal tibia including: simple comminuted, lateral wedge, depression, medial wedge, bicondylar, combi
January 13, 2016Medical DevicesFDA
Synthes (USA) Products LLC recalls SYNTHES; 13.5MM Medullary Reamer Head; 352.135 NON STERILE. Part of the Flexible Reamer Set for orthopedic surgical proc
Hazard: The 13.5mm Medullary Reamer head, lot #F-17180, is etched as dimension 13.5mm while the actual dimension is 14mm.
December 16, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Fem
Hazard: TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the handle. Or 2. Cause the bottom of the cannulated connecting screw to tighten against the top of the locking mechanism preventing advancement of the locking mechanism.
December 2, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls 2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, r
Hazard: It was discovered that the above part number and lot of 2.4mm Locking Screws may contain the incorrect package insert (GP2613, VA-LCP Distal Radius System). The correct package insert for the above part is GP2615 (Modular Mini Fragment LCP System).
November 25, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Synthes Application Instrument for Sternal ZIPFIX. Indications for use include primary or secondary closure/repair of
Hazard: Certain lots of the Application Instrument for Sternal ZIPFIX that were previously recalled on August 19, 2015 and subsequently repaired and returned to customers between August 26, 2015 and September 23, 2015. The screws on the repaired instruments have the potential to break or loosen. DePuy Synthes is requesting the return of the repaired device in exchange for a new device.
November 18, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Synthes Small Notch Titanium Reconstructive Plate set (SNTRP). Ti PLATE/ROD/PLATE 231mm 6HOLES/5.0mm ROD/17HOLES. Orth
Hazard: The Synthes SNTRP set was released for sale as a Trauma product in 1996 with specific trauma indications and contraindicated for certain spine applications. In 1999, a module was released containing a small offering of the SNTRP plates. Synthes determined that the product in the SNTRP module is being used in spine surgeries, which are a contraindicated use.
November 4, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Synthes Threaded Plate Holder- Long for 3.5mm Locking Hold; Part Numbers 324.031. The Threaded Plate Holder can be
Hazard: The affected part number and lot listed above is incorrectly etched with the wrong part number and may not fit into a Small Fragment Plate.
November 4, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls 2.7 mm Locking Screw SLF-TPNG with T8 Stardrive Recess 12mm. The Synthes modular Mini Fragment LCP System is intended f
Hazard: Specific lots of the SS 2.7mm locking screw have been mislabeled as SS 2.4mm locking screw.
October 28, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls 5.0mm Unit Rod 270mm, 5.0mm Unit Rod 290mm, 5.0mm Unit Rod 310mm, 5.0mm Unit Rod 330mm, 5.0mm Unit Rod 350mm, 5.0mm Un
Hazard: This product was produced using a finishing process not identified as part of the manufacturing specification. The process used with the lots subject to this Recall was a bead blast process. (Bead Blasting vs. Shot Peened).
October 21, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Synthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head
Hazard: Certain lots of the Screw Inserter may be etched with incorrect graphics. The orientation of the etched image of the oblique angle on the Screw Inserter may not correspond to the tip of the angle of the instrument.
October 21, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Inserter for Titanium Elastic Nails (TEN), orthopedic device
Hazard: potential for mechanical failures such as breakage.
October 21, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Synthes 2.7MM/3.5MM VA-LCP Medial Distal Humerus Plates 10H/RT/186mm- Sterile and Synthes 2.7MM/3.5MM VA-LCP Extended Me
Hazard: The 2.7MM/3.5MM VA-LCP Humerus Plates were found to be labeled incorrectly. Part number 02.117.410S was incorrectly etched with part number 02.117.610S
October 14, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Synthes Periarticular Insertion Handle for 4.5mm LCP Proximal Tibia Plates and Periarticular Aiming Arm for 4.5mm LCP Pr
Hazard: Affected lots of Periarticular Aiming Arms and Insertion Handles (Part Numbers 03.120.001, 03.120.002, 03.120.004 and 03.120.005) for the 4.5 mm LCP Proximal Tibia Plates are not able to be secured to the subsequent versions of these devices as intended.
September 30, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is in
Hazard: The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.
September 30, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls Dynamic Locking Screw System (DLS) 3.7mm & Dynamic Locking Screw (DLS) 5.0mm; intended for use in long bone fractures, t
Hazard: DLS pin breakage during planned implant removal, after uneventful and successful healing of the fracture. This notice to additional consignees identified after recall Z-0390-2014, related to complaints of breakage located at the bottom of the screws and under the locking screw head.
September 9, 2015Medical DevicesFDA
Synthes (USA) Products LLC recalls VEPTR Vertical Expandable Prosthetic Titanium Rib, inferior cradles also known as caudal rib supports. The VEPTR/VEPTR
Hazard: The last hole on certain VEPTR Inferior Cradles may be partially formed. Fully lengthening the construct to the partially formed last hole may result in mechanical failure of the construct due to rib sleeve breakage, and may require surgury to exchange components. Effects may also include pain, loss of deformity correction, and/or soft tissue/visceral injury.